Back to search
FDAFDA · Food and drugs·Recall number Z-1295-2021

VITROS Chemistry Products Calibrator Kit 4-(Calibrator values for TIBC only). Used to calibrate VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems Product Code: 1204668 Impacted Products: A. VITROS Chemistry Products TIBC Kit Product No: 7541 (UDI 8750005178) All current and new lots B. VITROS Chemistry Products Fe Slides Product No: 1924547 90 Slides (UDI 10758750000289) Product No: 1515808 300 slides (UDI 10758750009244) Product No: 6802198 (Japan) 90 Slides (UDI 10758750002276) All current, new and previous lots

Published March 31, 2021 · Recall initiated: February 18, 2021

Summary

FDA records medical device Class II recall Z-1295-2021, published March 31, 2021 and initiated February 18, 2021. Recalling firm: Ortho-Clinical Diagnostics, Inc.; Status: Terminated; Reported quantity: 7546 US and 15280 OUS kits; Lot: Lot Numbers: 0449 (exp. 28-Aug-2021) 0459 (exp. 10-Feb-2022) 0460 (exp. 01-Oct-2022) UDI: (10758750008537). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential total iron-binding capacity (TIBC) results. The VITROS TIBC performance shift will start with the newest Generation (GEN) of VITROS Fe Slides (GEN 26), the potential shift in TIBC patient values of GEN 26 and above will be approximately 80 ug/dL lower across the measuring range than other available GENS of VITROS Fe Slides (GENs 22-25).

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Ortho-Clinical Diagnostics, Inc.
Distributor
Worldwide distribution - US Nationwide distribution and the countries of Australia, Belgium, Brazil, Canada L3R 4G5, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Norway, Poland, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom.
Lot
Lot Numbers: 0449 (exp. 28-Aug-2021) 0459 (exp. 10-Feb-2022) 0460 (exp. 01-Oct-2022) UDI: (10758750008537)
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was VITROS Chemistry Products Calibrator Kit 4-(Calibrator values for TIBC only). Used to calibrate VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and VITROS 5600/XT 7600 Integrated Systems Product Code: 1204668 Impacted Products: A. VITROS Chemistry Products TIBC Kit Product No: 7541 (UDI 8750005178) All current and new lots B. VITROS Chemistry Products Fe Slides Product No: 1924547 90 Slides (UDI 10758750000289) Product No: 1515808 300 slides (UDI 10758750009244) Product No: 6802198 (Japan) 90 Slides (UDI 10758750002276) All current, new and previous lots recalled?

Yes. FDA published recall Z-1295-2021 on March 31, 2021. Reported hazard: Potential total iron-binding capacity (TIBC) results. The VITROS TIBC performance shift will start with the newest Generation (GEN) of VITROS Fe Slides (GEN 26), the potential shift in TIBC patient values of GEN 26 and above will be approximately 80 ug/dL lower across the measuring range than other available GENS of VITROS Fe Slides (GENs 22-25).

What should I do if I have this product?

Voluntary: Firm initiated · Letter