NautilusTM ECMO Oxygenator with Balance Biosurface-used in extracorporeal life support procedures to oxygenate blood, remove carbon dioxide and regulate blood temperature. Product Number: (1) 48145; (2) 48145E

Published March 27, 2024 · Recall initiated: October 11, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

May have increased water path restriction within the heat exchanger, leading to degraded heat exchange performance during use. The reduction of water flow and reduced heat exchange performance may be experienced immediately upon priming or may worsen over the duration of use. The potential harm as stated in the IFU includes hypothermia due to inadequate heat transfer

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
Mc3 Inc
Distributor
US Nationwide distribution.
Lot
GTIN: (1) 10854916006888; (2) 10854916006901,20854916006908 (New GTIN (USA only)After reboxing). All Serial Numbers between 481022923 and 481024039
Risk classification
Class II
Area (FDA)
device
Country
United States