cobas¿ 4800 system z480 instrument. Integrates fully automated total nucleic acid isolation directly from secondary sample tubes.

Published May 29, 2013 · Recall initiated: April 1, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

In rare instances, channel shifted results have been generated with cobas¿ 4800 assays run on v1.1.1 of the cobas¿ 4800 system. Raw data review indicates that, in these cases, all signals are shifted by one channel: channel 1 results report as channel 2, channel 2 results report as channel 3, channel 3 results report as channel 4, and channel 4 results report as channel 1. In the reported

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Roche Molecular Systems, Inc.
Distributor
Nationwide Distribution.
Lot
Material Number 05200881001 - all lot/serial numbers.
Risk classification
Class II
Area (FDA)
device
Country
United States