Ellipse, Tiered-therapy Cardioverter/Defibrillator, REF (UDI/GTIN): CD1377-36QC (05414734507646), CD2411-36C (05414734507585), CD1277-36 (05414734506045), CD1377-36C (05414734507622), CD2277-36Q (05414734505932), CD2377-36QC (05414734507523), with Merlin Patient Care System with software model 3330
Published March 4, 2020 · Recall initiated: January 22, 2020
Summary
FDA records medical device Class II recall Z-1351-2020, published March 4, 2020 and initiated January 22, 2020. Recalling firm: St Jude Medical, Cardiac Rhythm Management Division; Status: Terminated; Reported quantity: 256; Lot: Serial #: 1138400, 1138406, 1138391, 1138393, 1138403, 1138366, 1138387, 1138396, 1138386, 1138399, 1138374, 1138362, 1138368, 1138397, 1138401, 1138372, 1138375, 1138382, 1138380, 1138405, 1138377, 1138371, 1138390, 1138398, 1138367, 1138395, 1138389, 1138379, 1138370, 1138388, 1138364, 1138373, 1138385, 1138394, 1138381, 1138404, 1138392, 1138384, 1138376, 1138369, 1138365, 1138465, 1138620, 1138573, 1138590, 1138407, 1138408, 1138409, 1138411, 1138412, 1138413, 1138415, 1138416, 1138417, 1138419, 1138421, 1138422, 1138423, 1138424, 1138429, 1138431, 1138435, 1138436, 1138437, 1138441, 1138442, 1138467, 1138468, 1138469, 1138513, 1138612, 1138613, 1138615, 1138616, 1138617, 1138624, 1138626, 1138420, 1138587, 1138628, 1138510, 1138619, 1138428, 1138432, 1138433, 1138445, 1138446, 1138447, 1138449, 1138451, 1138455, 1138466, 1138470, 1138471, 1138474, 1138479, 1138488, 1138502, 1138507, 1138523, 1138538, 1138544, 1138548, 1138556, 1138575, 1138576, 1138577, 1138580, 1138597, 1138599, 1138608, 1138495, 1138503, 1138547, 1138563, 1138565, 1138581, 1138585, 1138543, 1138461, 1138501, 1138505, 1138508, 1138545, 1138546, 1138553, 1138557, 1138566, 1138567, 1138583, 1138588, 1138486, 1138489, 1138490, 1138491, 1138496, 1138497, 1138500, 1138542, 1138549, 1138550, 1138551, 1138555, 1138559, 1138560, 1138561, 1138564, 1138570, 1138579, 1138582, 1138586, 1138611, 1138618, 1138472, 1138482, 1138595, 1138589, 1138614, 1138434, 1138440, 1138443, 1138448, 1138450, 1138452, 1138453, 1138456, 1138457, 1138458, 1138459, 1138460, 1138462, 1138463, 1138464, 1138473, 1138476, 1138478, 1138480, 1138481, 1138514, 1138515, 1138516, 1138518, 1138519, 1138520, 1138521, 1138522, 1138524, 1138525, 1138526, 1138527, 1138528, 1138529, 1138530, 1138531, 1138532, 1138533, 1138534, 1138536, 1138537, 1138539, 1138540, 1138594, 1138596, 1138600, 1138601, 1138602, 1138603, 1138604, 1138605, 1138606, 1138607, 1138609, 1138622, 1138623, 1138631, 1138632, 1138633, 1138634, 1138635, 1138636, 1138487, 1138494, 1138552, 1138554, 1138610, 1138621, 1138493, 1138499, 1138504, 1138569, 1138627, 1138484, 1138571, 1138425, 1138426, 1138438, 1138439, 1138444, 1138498, 1138492, 1138509, 1138541, 1138558, 1138562, 1138568, 1138584, 1138578, 1138511, 1138512, 1138572, 1138591, 1138592, 1138593, 1138625, 1138485, 1138574, 1138430, 1138598, 1138402, 1138363, 1138454, 1138410, 1138383, 1138378, 1138517, 1138535, 1138418, 1138506, 1138414, 1138427, 1138475, 1138477, 1138483, 1138630. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Past updates to programmers and transmitters may lead some implantable cardioverter defibrillators (ICD) to lose wireless radiofrequency (RF) communication; which means patients can no longer be interrogated with wireless RF telemetry or monitored remotely.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- St Jude Medical, Cardiac Rhythm Management Division
- Distributor
- U.S.: WV, MS, TX, DE, OH, AZ, MI, NC, KY, WA, NY, CA, IN, MO, MD, NJ, FL, PA, OK. O.U.S.: United Kingdom, Turkey, Sweden, Spain, Portugal, Poland, Philippines, Pakistan, Norway, New Zealand, Netherlands, Japan, Italy, Hong Kong, Greece, Germany, France, Finland, Denmark, Canada, Belgium, Austria, Australia
- Lot
- Serial #: 1138400, 1138406, 1138391, 1138393, 1138403, 1138366, 1138387, 1138396, 1138386, 1138399, 1138374, 1138362, 1138368, 1138397, 1138401, 1138372, 1138375, 1138382, 1138380, 1138405, 1138377, 1138371, 1138390, 1138398, 1138367, 1138395, 1138389, 1138379, 1138370, 1138388, 1138364, 1138373, 1138385, 1138394, 1138381, 1138404, 1138392, 1138384, 1138376, 1138369, 1138365, 1138465, 1138620, 1138573, 1138590, 1138407, 1138408, 1138409, 1138411, 1138412, 1138413, 1138415, 1138416, 1138417, 1138419, 1138421, 1138422, 1138423, 1138424, 1138429, 1138431, 1138435, 1138436, 1138437, 1138441, 1138442, 1138467, 1138468, 1138469, 1138513, 1138612, 1138613, 1138615, 1138616, 1138617, 1138624, 1138626, 1138420, 1138587, 1138628, 1138510, 1138619, 1138428, 1138432, 1138433, 1138445, 1138446, 1138447, 1138449, 1138451, 1138455, 1138466, 1138470, 1138471, 1138474, 1138479, 1138488, 1138502, 1138507, 1138523, 1138538, 1138544, 1138548, 1138556, 1138575, 1138576, 1138577, 1138580, 1138597, 1138599, 1138608, 1138495, 1138503, 1138547, 1138563, 1138565, 1138581, 1138585, 1138543, 1138461, 1138501, 1138505, 1138508, 1138545, 1138546, 1138553, 1138557, 1138566, 1138567, 1138583, 1138588, 1138486, 1138489, 1138490, 1138491, 1138496, 1138497, 1138500, 1138542, 1138549, 1138550, 1138551, 1138555, 1138559, 1138560, 1138561, 1138564, 1138570, 1138579, 1138582, 1138586, 1138611, 1138618, 1138472, 1138482, 1138595, 1138589, 1138614, 1138434, 1138440, 1138443, 1138448, 1138450, 1138452, 1138453, 1138456, 1138457, 1138458, 1138459, 1138460, 1138462, 1138463, 1138464, 1138473, 1138476, 1138478, 1138480, 1138481, 1138514, 1138515, 1138516, 1138518, 1138519, 1138520, 1138521, 1138522, 1138524, 1138525, 1138526, 1138527, 1138528, 1138529, 1138530, 1138531, 1138532, 1138533, 1138534, 1138536, 1138537, 1138539, 1138540, 1138594, 1138596, 1138600, 1138601, 1138602, 1138603, 1138604, 1138605, 1138606, 1138607, 1138609, 1138622, 1138623, 1138631, 1138632, 1138633, 1138634, 1138635, 1138636, 1138487, 1138494, 1138552, 1138554, 1138610, 1138621, 1138493, 1138499, 1138504, 1138569, 1138627, 1138484, 1138571, 1138425, 1138426, 1138438, 1138439, 1138444, 1138498, 1138492, 1138509, 1138541, 1138558, 1138562, 1138568, 1138584, 1138578, 1138511, 1138512, 1138572, 1138591, 1138592, 1138593, 1138625, 1138485, 1138574, 1138430, 1138598, 1138402, 1138363, 1138454, 1138410, 1138383, 1138378, 1138517, 1138535, 1138418, 1138506, 1138414, 1138427, 1138475, 1138477, 1138483, 1138630
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATHP Hip Plating System, Item Nos. 14-451070 14-451075 14-451080 14-451085 14-451090 14-451095 14-451100March 4, 2020
- FDAProtective Cover Vivo 55/65 (Used for Vivo 50/65), Part No. 006344. It is an accessory (i.e. protective cover)March 4, 2020
- FDAPhilips DigiTrak XT Holter Recorder Software version number B.1 and C.2. The DigiTrak XT Recorder is availableMarch 4, 2020
- FDAZimmer Sterile Disposable Tourniquet Cuff with Protective Sleeve and PLC, Dual Port, Dual Bladder, Item No.March 4, 2020
- FDASelenia Dimensions Mammography System, Model number ASY-04160, Selenia Dimensions Gantry, UDI 15420045505636,March 4, 2020
- FDAMedline ROBOTIC-LF REF CDS984543B Lot 20AKA927 Expiration 2020-09-30 GTIN (01)10888277056350March 4, 2020
Questions about this recall
Was Ellipse, Tiered-therapy Cardioverter/Defibrillator, REF (UDI/GTIN): CD1377-36QC (05414734507646), CD2411-36C (05414734507585), CD1277-36 (05414734506045), CD1377-36C (05414734507622), CD2277-36Q (05414734505932), CD2377-36QC (05414734507523), with Merlin Patient Care System with software model 3330 recalled?
Yes. FDA published recall Z-1351-2020 on March 4, 2020. Reported hazard: Past updates to programmers and transmitters may lead some implantable cardioverter defibrillators (ICD) to lose wireless radiofrequency (RF) communication; which means patients can no longer be interrogated with wireless RF telemetry or monitored remotely.
What should I do if I have this product?
Voluntary: Firm initiated · Letter