NeuRx Diaphragm Pacing System (NeuRx DPS), Affected part numbers: Part # Part Name 20-0035 NeuRx DPS System Kit 20-0036 NeuRx Diaphragm Pacing System, Patient Kit, SCI 20-0028 NeuRx DPS Surgical Kit, HDE 20-0040 NeuRx Diaphragm Pacing System Spare Sterile Components 21-0001-99 Mapping Probe 21-0004-99 Indifferent Electrode 21-0007-99 Lead Tunneler Set 21-0009-99 Blue Cable Set 21-0010-99 Transducer Tubing 21-0017-99 PermaLoc Electrode 21-0023-99 Green Cable 21-0024-99 Orange Cable 21-0025-99 Yellow Cable 21-0026-99 White Cable 77-0050 Surgeon Instruction Manual 20-0045 NeuRx DPS System Kit, ALS HDE 20-0046 NeuRx Diaphragm Pacing System, Patient Kit, ALS Marketed under HDE H070003 as well as H100006
Published July 27, 2022 · Recall initiated: May 31, 2022
Summary
FDA records medical device Class III recall Z-1393-2022, published July 27, 2022 and initiated May 31, 2022. Recalling firm: Synapse Biomedical Inc; Status: Terminated; Reported quantity: 71; Lot: UDI-DI: 00852184003007 00852184003045 00852184003038 00852184003267 00852184003090 00852184003106 00852184003113 00852184003120 00852184003137 00852184003144 00852184003151 00852184003168 00852184003175 00852184003182 00852184003281 00852184003052 Lots/Serial Numbers: 20-0035-113021-3-3 20-0035-120221-2-1 20-0035-121621-5-2 20-0035-121621-5-3 20-0035-010722-5-4 20-0035-010722-5-5 20-0035-030722-2-2 20-0035-031122-5-1 20-0035-031122-5-2 20-0035-031122-5-5 20-0035-033122-5-1 20-0035-033122-5-2 20-0035-033122-5-3 20-0035-040622-4-1 20-0035-040622-4-2 20-0035-040622-4-3 20-0035-040622-4-2 20-0035-042122-1-1 20-0028-060221-5 20-0028-062421-5 20-0028-062521-5 20-0028-072821-5 20-0028-031022-5 20-0040-012221-10 20-0040-012221-10 20-0040-021921-9 20-0040-012221-10 20-0040-012221-10 20-0040-021921-9 20-0040-021921-9 20-0040-021921-9 20-0040-021921-9 20-0040-012621-10 20-0040-012621-10 20-0040-012621-10 20-0040-012621-10 20-0040-060221-5 20-0040-060221-5 20-0040-060221-5 20-0040-060221-5 20-0040-060221-5 20-0040-060821-5 20-0040-060821-5 20-0040-060821-5 20-0040-060821-5 20-0040-060821-5 20-0040-071321-6 20-0040-071321-6 20-0040-092021-10 20-0040-092021-10 20-0040-071521-2 20-0040-092021-10 20-0040-092021-10 20-0040-092021-10 20-0040-100721-10 20-0040-100721-10 20-0040-100721-10 20-0040-100721-10 20-040-100721-10 20-0040-100721-10 20-0040-100721-10 20-0040-122121-10 20-0040-122121-10 20-0040-122121-10 20-0040-122121-10 20-0040-010622-10 20-0040-121621-1 20-0040-020822-10 20-0040-010622-10 20-0040-040122-2. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Firm has revised labeling to consistently call out storage conditions and revised the instructions for use to clarify that the temperature range symbol is intended to indicate storage conditions (versus transport conditions or tolerance conditions for the package material).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Synapse Biomedical Inc
- Distributor
- Domestic distribution to AL CA CO FL GA IL MA MI MN NC NJ NY OH OK OR PA SC TX WV. Foreign distribution pending.
- Lot
- UDI-DI: 00852184003007 00852184003045 00852184003038 00852184003267 00852184003090 00852184003106 00852184003113 00852184003120 00852184003137 00852184003144 00852184003151 00852184003168 00852184003175 00852184003182 00852184003281 00852184003052 Lots/Serial Numbers: 20-0035-113021-3-3 20-0035-120221-2-1 20-0035-121621-5-2 20-0035-121621-5-3 20-0035-010722-5-4 20-0035-010722-5-5 20-0035-030722-2-2 20-0035-031122-5-1 20-0035-031122-5-2 20-0035-031122-5-5 20-0035-033122-5-1 20-0035-033122-5-2 20-0035-033122-5-3 20-0035-040622-4-1 20-0035-040622-4-2 20-0035-040622-4-3 20-0035-040622-4-2 20-0035-042122-1-1 20-0028-060221-5 20-0028-062421-5 20-0028-062521-5 20-0028-072821-5 20-0028-031022-5 20-0040-012221-10 20-0040-012221-10 20-0040-021921-9 20-0040-012221-10 20-0040-012221-10 20-0040-021921-9 20-0040-021921-9 20-0040-021921-9 20-0040-021921-9 20-0040-012621-10 20-0040-012621-10 20-0040-012621-10 20-0040-012621-10 20-0040-060221-5 20-0040-060221-5 20-0040-060221-5 20-0040-060221-5 20-0040-060221-5 20-0040-060821-5 20-0040-060821-5 20-0040-060821-5 20-0040-060821-5 20-0040-060821-5 20-0040-071321-6 20-0040-071321-6 20-0040-092021-10 20-0040-092021-10 20-0040-071521-2 20-0040-092021-10 20-0040-092021-10 20-0040-092021-10 20-0040-100721-10 20-0040-100721-10 20-0040-100721-10 20-0040-100721-10 20-040-100721-10 20-0040-100721-10 20-0040-100721-10 20-0040-122121-10 20-0040-122121-10 20-0040-122121-10 20-0040-122121-10 20-0040-010622-10 20-0040-121621-1 20-0040-020822-10 20-0040-010622-10 20-0040-040122-2
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAENDURO SPECIAL KEY TIB.LOCK.RING F1/10MM product code NP462R used with EnduRo Total Knee System. instrumentJuly 27, 2022
- FDAMODULIFT VBR SZ.M 4MM SET SCREW DRIVER, MF790R, part of Modulift VBR Medium Instrument Set, ST-0441July 27, 2022
- FDATargil's St. Amand's Chicken Breading packaged in 25lb boxesJuly 27, 2022
- FDAATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparationJuly 27, 2022
- FDATargil's Parker's Flavor Flour packaged in 25lb boxesJuly 27, 2022
- FDASoluble Transferrin Receptors (STFR), Catalog Number TF10159 (Research Use Only). IVDJuly 27, 2022
Questions about this recall
Was NeuRx Diaphragm Pacing System (NeuRx DPS), Affected part numbers: Part # Part Name 20-0035 NeuRx DPS System Kit 20-0036 NeuRx Diaphragm Pacing System, Patient Kit, SCI 20-0028 NeuRx DPS Surgical Kit, HDE 20-0040 NeuRx Diaphragm Pacing System Spare Sterile Components 21-0001-99 Mapping Probe 21-0004-99 Indifferent Electrode 21-0007-99 Lead Tunneler Set 21-0009-99 Blue Cable Set 21-0010-99 Transducer Tubing 21-0017-99 PermaLoc Electrode 21-0023-99 Green Cable 21-0024-99 Orange Cable 21-0025-99 Yellow Cable 21-0026-99 White Cable 77-0050 Surgeon Instruction Manual 20-0045 NeuRx DPS System Kit, ALS HDE 20-0046 NeuRx Diaphragm Pacing System, Patient Kit, ALS Marketed under HDE H070003 as well as H100006 recalled?
Yes. FDA published recall Z-1393-2022 on July 27, 2022. Reported hazard: Firm has revised labeling to consistently call out storage conditions and revised the instructions for use to clarify that the temperature range symbol is intended to indicate storage conditions (versus transport conditions or tolerance conditions for the package material).
What should I do if I have this product?
Voluntary: Firm initiated · Letter