Soluble Transferrin Receptors (STFR), Catalog Number TF10159 (Research Use Only). IVD
Published July 27, 2022 · Recall initiated: June 9, 2022
Summary
FDA records medical device Class III recall Z-1422-2022, published July 27, 2022 and initiated June 9, 2022. Recalling firm: Randox Laboratories Ltd.; Status: Terminated; Reported quantity: 12 kits (US); Lot: UDI-DI (GTIN): 05055273215564 All lots within expiry. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration curves that can results in non-numerical results. This affects Siemens Atellica CH analysers only. Customer have reported an atypical calibration curve that results in the values around the low QC range to produce a non-numerical result. It appears to be related to the bottle values of the calibrator producing a curve shape that is too linear for the logitlog calculation to calculate. This issue can only be identified in the Atellica graphical view of the calibration. In most of the cases when this has occurred, it is picked up by the QC generating a Math Error, however, there are reports of QC shifts after calibration without the Math Error flagged that may be related to this issue. Results reported with STFR Level 1, with a concentration less than 1.7mg/L, using TF10159 and TF10161 can be underestimated by up to 30%. This issue has been reported with reagent lots 575475 and 562691, calibrator lots 537725 and 562696. No other complaints received.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Randox Laboratories Ltd.
- Distributor
- US Nationwide distribution in the state of IL.
- Lot
- UDI-DI (GTIN): 05055273215564 All lots within expiry
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- Ireland
Related recalls
- FDARX Series CYSC- Cystatin C Reagent -Intended for in vitro diagnostic (IVD) use as Randox Cystatin C assay onJuly 31, 2024
- FDAClinical Chemistry Calibration Serum (Catalog number Cal2351) that is 20x5ml lyophilized human based serumMay 15, 2024
- FDAEvidence MultiSTAT with software v 3.7-Analyzer intended for the qualitative determination of parent drugMay 15, 2024
- FDAMicroalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567April 24, 2024
- FDAATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparationJuly 27, 2022
- FDASensis with software version VD12A. A diagnostic and administrative tool supporting hemodynamicJuly 27, 2022
Questions about this recall
Was Soluble Transferrin Receptors (STFR), Catalog Number TF10159 (Research Use Only). IVD recalled?
Yes. FDA published recall Z-1422-2022 on July 27, 2022. Reported hazard: A customer reported a problem with the reagent and calibrator. Randox Laboratories carried out an investigation and confirmed the issue. The complaint is pertaining to atypical calibration curves that can results in non-numerical results. This affects Siemens Atellica CH analysers only. Customer have reported an atypical calibration curve that results in the values around the low QC range to produce a non-numerical result. It appears to be related to the bottle values of the calibrator producing a curve shape that is too linear for the logitlog calculation to calculate. This issue can only be identified in the Atellica graphical view of the calibration. In most of the cases when this has occurred, it is picked up by the QC generating a Math Error, however, there are reports of QC shifts after calibration without the Math Error flagged that may be related to this issue. Results reported with STFR Level 1, with a concentration less than 1.7mg/L, using TF10159 and TF10161 can be underestimated by up to 30%. This issue has been reported with reagent lots 575475 and 562691, calibrator lots 537725 and 562696. No other complaints received.
What should I do if I have this product?
Voluntary: Firm initiated · Letter