Spacelabs Pediatric Flow Sensor Kit, PN: 376-0561-00. This kit is for Spacelabs Healthcare Blease 700/900 Series Ventilators. Designed specifically for the mechanical ventilation of adult and pediatric patients under general anesthesia.
Published April 29, 2015 · Recall initiated: March 5, 2015
Summary
FDA records medical device Class II recall Z-1458-2015, published April 29, 2015 and initiated March 5, 2015. Recalling firm: Del Mar Reynolds Medical, Ltd.; Status: Terminated; Reported quantity: 1040 units total (398 in the US and 642 international); Lot: PN: 376-0561-00. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Reports of inaccurate low flow readings. Monitored inspiratory tidal volume (VTi) and expiratory tidal volume (Vte) measurements from the pediatric flow sensor are reporting out of specification low compared to the actual delivered volumes being administered to the patient.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Del Mar Reynolds Medical, Ltd.
- Distributor
- Worldwide Distribution-US including the states of Louisiana, North Carolina, and Washington and the countries of Argentina, Brazil, China, Colombia, Ecuador, Finland, Great Britain, India, Libya, Mexico, Nicaragua, Oman, Philippines, Saudi Arabia, Slovakia, Sri Lanka, United Arab Emirates, and Vietnam.
- Lot
- PN: 376-0561-00.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United Kingdom
Related recalls
- FDAArkon Anesthesia Delivery System, Model 99999. Product Usage: Continuous Ventilator The Arkon AnesthesiaMarch 22, 2017
- FDABorden Good 2 Gether, Mild Cheddar Cheese & Pretzels, UPC: 5300001111, Net Wt 2 0z (57 g), Dist. By DFA DairyApril 29, 2015
- FDATorque Limiting Handle (Aspen/ Alpine) Model Number 6200-1109 Product Usage: Provides posterior fixation inApril 29, 2015
- FDASpaceLabs Arkon Anesthesia Delivery System with Version 2.0 Software installed. This system includes aApril 23, 2014
- FDACAS I/II Absorbers that are parts of the BleaseSirius Anesthesia Workstation, the BleaseFocus AnesthesiaNovember 27, 2013
- FDASpacelabs ARKON Anesthesia Workstation. Spacelabs ARKON Anesthesia Workstation may be used for the delivery ofJuly 31, 2013
Questions about this recall
Was Spacelabs Pediatric Flow Sensor Kit, PN: 376-0561-00. This kit is for Spacelabs Healthcare Blease 700/900 Series Ventilators. Designed specifically for the mechanical ventilation of adult and pediatric patients under general anesthesia. recalled?
Yes. FDA published recall Z-1458-2015 on April 29, 2015. Reported hazard: Reports of inaccurate low flow readings. Monitored inspiratory tidal volume (VTi) and expiratory tidal volume (Vte) measurements from the pediatric flow sensor are reporting out of specification low compared to the actual delivered volumes being administered to the patient.
What should I do if I have this product?
Voluntary: Firm initiated · Letter