Spacelabs Pediatric Flow Sensor Kit, PN: 376-0561-00. This kit is for Spacelabs Healthcare Blease 700/900 Series Ventilators. Designed specifically for the mechanical ventilation of adult and pediatric patients under general anesthesia.

Published April 29, 2015 · Recall initiated: March 5, 2015

Summary

FDA records medical device Class II recall Z-1458-2015, published April 29, 2015 and initiated March 5, 2015. Recalling firm: Del Mar Reynolds Medical, Ltd.; Status: Terminated; Reported quantity: 1040 units total (398 in the US and 642 international); Lot: PN: 376-0561-00. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Reports of inaccurate low flow readings. Monitored inspiratory tidal volume (VTi) and expiratory tidal volume (Vte) measurements from the pediatric flow sensor are reporting out of specification low compared to the actual delivered volumes being administered to the patient.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Del Mar Reynolds Medical, Ltd.
Distributor
Worldwide Distribution-US including the states of Louisiana, North Carolina, and Washington and the countries of Argentina, Brazil, China, Colombia, Ecuador, Finland, Great Britain, India, Libya, Mexico, Nicaragua, Oman, Philippines, Saudi Arabia, Slovakia, Sri Lanka, United Arab Emirates, and Vietnam.
Lot
PN: 376-0561-00.
Risk classification
Class II
Area (FDA)
device
Country
United Kingdom

Related recalls

Questions about this recall

Was Spacelabs Pediatric Flow Sensor Kit, PN: 376-0561-00. This kit is for Spacelabs Healthcare Blease 700/900 Series Ventilators. Designed specifically for the mechanical ventilation of adult and pediatric patients under general anesthesia. recalled?

Yes. FDA published recall Z-1458-2015 on April 29, 2015. Reported hazard: Reports of inaccurate low flow readings. Monitored inspiratory tidal volume (VTi) and expiratory tidal volume (Vte) measurements from the pediatric flow sensor are reporting out of specification low compared to the actual delivered volumes being administered to the patient.

What should I do if I have this product?

Voluntary: Firm initiated · Letter