Spacelabs ARKON Anesthesia Workstation. Spacelabs ARKON Anesthesia Workstation may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer.
Published July 31, 2013 · Recall initiated: June 7, 2013
Summary
FDA records medical device Class II recall Z-1771-2013, published July 31, 2013 and initiated June 7, 2013. Recalling firm: Del Mar Reynolds Medical, Ltd.; Status: Terminated; Reported quantity: 9 units (4 in the US and 5 outside US); Lot: ARKN-000001, ARKN-000005, ARKN-000010, ARKN-000015, ARKN-000016, ARKN-000017, ARKN-000018, ARKN-000020, ARKN-000021. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Spacelabs ARKON Anesthesia System is recalled due to software defect. The System Status Computer may incorrectly determine that a communication error has occurred. If this situation occurs, a buzzer will activate and a yellow triangle will be displayed on the System Status Computer screen that alerts the user of this error.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Del Mar Reynolds Medical, Ltd.
- Distributor
- Worldwide Distribution - USA including South Carolina and internationally to Canada and Great Britian.
- Lot
- ARKN-000001, ARKN-000005, ARKN-000010, ARKN-000015, ARKN-000016, ARKN-000017, ARKN-000018, ARKN-000020, ARKN-000021.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United Kingdom
Related recalls
- FDAArkon Anesthesia Delivery System, Model 99999. Product Usage: Continuous Ventilator The Arkon AnesthesiaMarch 22, 2017
- FDASpacelabs Pediatric Flow Sensor Kit, PN: 376-0561-00. This kit is for Spacelabs Healthcare Blease 700/900April 29, 2015
- FDASpaceLabs Arkon Anesthesia Delivery System with Version 2.0 Software installed. This system includes aApril 23, 2014
- FDACAS I/II Absorbers that are parts of the BleaseSirius Anesthesia Workstation, the BleaseFocus AnesthesiaNovember 27, 2013
- FDAAlboGraft Woven with Collagen, Straight 30x26 Catalog Number: ATC3026 AlboGraft Vascular Grafts are made ofJuly 31, 2013
- FDAEndopath Electrosurgery Probe Plus II, twelve (12) different devices, including hooks, spatulas, right angles,July 31, 2013
Questions about this recall
Was Spacelabs ARKON Anesthesia Workstation. Spacelabs ARKON Anesthesia Workstation may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. recalled?
Yes. FDA published recall Z-1771-2013 on July 31, 2013. Reported hazard: Spacelabs ARKON Anesthesia System is recalled due to software defect. The System Status Computer may incorrectly determine that a communication error has occurred. If this situation occurs, a buzzer will activate and a yellow triangle will be displayed on the System Status Computer screen that alerts the user of this error.
What should I do if I have this product?
Voluntary: Firm initiated · Letter