Spacelabs ARKON Anesthesia Workstation. Spacelabs ARKON Anesthesia Workstation may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer.

Published July 31, 2013 · Recall initiated: June 7, 2013

Summary

FDA records medical device Class II recall Z-1771-2013, published July 31, 2013 and initiated June 7, 2013. Recalling firm: Del Mar Reynolds Medical, Ltd.; Status: Terminated; Reported quantity: 9 units (4 in the US and 5 outside US); Lot: ARKN-000001, ARKN-000005, ARKN-000010, ARKN-000015, ARKN-000016, ARKN-000017, ARKN-000018, ARKN-000020, ARKN-000021. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Spacelabs ARKON Anesthesia System is recalled due to software defect. The System Status Computer may incorrectly determine that a communication error has occurred. If this situation occurs, a buzzer will activate and a yellow triangle will be displayed on the System Status Computer screen that alerts the user of this error.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Del Mar Reynolds Medical, Ltd.
Distributor
Worldwide Distribution - USA including South Carolina and internationally to Canada and Great Britian.
Lot
ARKN-000001, ARKN-000005, ARKN-000010, ARKN-000015, ARKN-000016, ARKN-000017, ARKN-000018, ARKN-000020, ARKN-000021.
Risk classification
Class II
Area (FDA)
device
Country
United Kingdom

Related recalls

Questions about this recall

Was Spacelabs ARKON Anesthesia Workstation. Spacelabs ARKON Anesthesia Workstation may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. recalled?

Yes. FDA published recall Z-1771-2013 on July 31, 2013. Reported hazard: Spacelabs ARKON Anesthesia System is recalled due to software defect. The System Status Computer may incorrectly determine that a communication error has occurred. If this situation occurs, a buzzer will activate and a yellow triangle will be displayed on the System Status Computer screen that alerts the user of this error.

What should I do if I have this product?

Voluntary: Firm initiated · Letter