FDAFDA · Food and drugs·Recall number Z-1460-2017

Arkon Anesthesia Delivery System, Model 99999. Product Usage: Continuous Ventilator The Arkon Anesthesia Delivery System is designed to provide emergency oxygen, vaporized agent delivery and manual ventilation in the event of a power failure scenario. Clinicians are able to use Emergency O2 and manually ventilate the patient, providing gas and agent. However, the anesthesiologist will not have access to mechanical ventilation or ventilator monitoring.

Published March 22, 2017 · Recall initiated: February 8, 2017

Summary

FDA records medical device Class II recall Z-1460-2017, published March 22, 2017 and initiated February 8, 2017. Recalling firm: Del Mar Reynolds Medical, Ltd.; Status: Terminated; Reported quantity: 325 devices sold to 35 US consignees and 76 devices sold to 14 international consignees; Lot: US Serial numbers: ARKN-000003, ARKN-000004, ARKN-000011, ARKN-000012, ARKN-000013, ARKN-000014, ARKN-000016, ARKN-000017, ARKN-000019, ARKN-000020, ARKN-000021, ARKN-000022, ARKN-000023, ARKN-000024, ARKN-000025, ARKN-000026, ARKN-000027, ARKN-000028, ARKN-000029, ARKN-000030, ARKN-000031, ARKN-000037, ARKN-000038, ARKN-000039, ARKN-000040, ARKN-000041, ARKN-000042, ARKN-000044, ARKN-000045, ARKN-000046, ARKN-000047, ARKN-000048, ARKN-000049, ARKN-000051, ARKN-000054, ARKN-000055, ARKN-000056, ARKN-000057, ARKN-000058, ARKN-000059, ARKN-000060, ARKN-000061, ARKN-000062, ARKN-000063, ARKN-000064, ARKN-000065, ARKN-000066, ARKN-000067, ARKN-000068, ARKN-000069, ARKN-000070, ARKN-000071, ARKN-000072, ARKN-000073, ARKN-000074, ARKN-000075, ARKN-000076, ARKN-000077, ARKN-000078, ARKN-000082, ARKN-000091, ARKN-000092, ARKN-000093, ARKN-000094, ARKN-000095, ARKN-000096, ARKN-000097, ARKN-000098, ARKN-000099, ARKN-000100, ARKN-000101, ARKN-000102, ARKN-000103, ARKN-000104, ARKN-000105, ARKN-000106, ARKN-000107, ARKN-000108, ARKN-000109, ARKN-000110, ARKN-000111, ARKN-000112, ARKN-000113, ARKN-000114, ARKN-000115, ARKN-000116, ARKN-000117, ARKN-000119, ARKN-000120, ARKN-000123, ARKN-000129, ARKN-000130, ARKN-000132, ARKN-000133, ARKN-000134, ARKN-000135, ARKN-000136, ARKN-000137, ARKN-000138, ARKN-000139, ARKN-000140, ARKN-000141, ARKN-000142, ARKN-000143, ARKN-000144, ARKN-000145, ARKN-000149, ARKN-000150, ARKN-000151, ARKN-000152, ARKN-000153, ARKN-000154, ARKN-000155, ARKN-000157, ARKN-000167, ARKN-000168, arkn-000169, arkn-000170, ARKN-000171, arkn-000173, arkn-000174, arkn-000175, arkn-000176, arkn-000177, arkn-000178, arkn-000179, arkn-000180, arkn-000181, arkn-000182, arkn-000183, arkn-000184, arkn-000185, ARKN-000186, ARKN-000187, ARKN-000188, ARKN-000189, ARKN-000190, ARKN-000191, ARKN-000192, ARKN-000193, ARKN-000194, ARKN-000196, ARKN-000197, ARKN-000198, ARKN-000199, ARKN-000200, ARKN-000201, ARKN-000202, ARKN-000203, ARKN-000204, ARKN-000205, ARKN-000206, ARKN-000207, ARKN-000208, ARKN-000209, ARKN-000210, ARKN-000211, ARKN-000212, ARKN-000213, ARKN-000214, ARKN-000215, ARKN-000216, ARKN-000217, ARKN-000220, ARKN-000221, ARKN-000222, arkn-000223, arkn-000224, arkn-000226, arkn-000227, arkn-000229, arkn-000230, arkn-000231, arkn-000232, arkn-000233, ARKN-000236, arkn-000237, arkn-000238, arkn-000239, arkn-000240, arkn-000241, arkn-000242, arkn-000243, arkn-000244, ARKN-000245, arkn-000248, arkn-000249, ARKN-000250, ARKN-000251, ARKN-000252, ARKN-000253, arkn-000254, arkn-000257, arkn-000258, ARKN-000259, ARKN-000260, arkn-000261, ARKN-000262, arkn-000263, arkn-000264, arkn-000266, arkn-000269, arkn-000270, arkn-000271, arkn-000272, arkn-000273, arkn-000274, arkn-000275, arkn-000276, arkn-000277, ARKN-000278, ARKN-000279, ARKN-000280, ARKN-000281, ARKN-000282, ARKN-000283, ARKN-000284, ARKN-000285, ARKN-000286, ARKN-000287, ARKN-000288, ARKN-000289, ARKN-000290, ARKN-000293, ARKN-000294, ARKN-000295, ARKN-000296, ARKN-000297, ARKN-000300, ARKN-000301, ARKN-000302, ARKN-000303, ARKN-000304, ARKN-000305, ARKN-000306, ARKN-000307, ARKN-000308, ARKN-000309, ARKN-000310, ARKN-000311, ARKN-000312, ARKN-000313, arkn-000314, arkn-000315, arkn-000316, arkn-000317, arkn-000318, ARKN-000319, arkn-000320, arkn-000321, arkn-000322, arkn-000323, arkn-000324, arkn-000325, arkn-000327, arkn-000328, arkn-000329, arkn-000330, ARKN-000332, ARKN-000333, ARKN-000334, ARKN-000335, ARKN-000336, ARKN-000337, ARKN-000338, ARKN-000339, ARKN-000340, ARKN-000341, ARKN-000342, ARKN-000343, ARKN-000344, ARKN-000345, ARKN-000353, ARKN-000360, ARKN-000361, ARKN-000362, ARKN-000363, ARKN-000364, ARKN-000365, ARKN-000366, ARKN-000367, ARKN-000369, ARKN-000370, ARKN-000371, ARKN-000372, ARKN-000373, ARKN-000374, ARKN-000376, ARKN-000377, ARKN-000378, ARKN-000379, ARKN-000380, ARKN-000381, ARKN-000382, ARKN-000383, ARKN-000384, ARKN-000385, ARKN-000386, ARKN-000387, ARKN-000388, ARKN-000389, ARKN-000390, ARKN-000391, ARKN-000392, ARKN-000393, ARKN-000394, . Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Spacelabs has received reports of the Arkon Anesthesia Delivery System, Model 99999, shutting down without warning. The power supply has failed in a way that does not trigger the normal on battery alarms.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Del Mar Reynolds Medical, Ltd.
Distributor
Worldwide Distribution - US nationwide and internationally in the following countries: BOLIVIA, CANADA, CHINA, ECUADOR, FRANCE, IRELAND, MALAYSIA, MEXICO, MOROCCO, NETHERLANDS, PANAMA, PERU, POLAND, UNITED KINGDOM.
Lot
US Serial numbers: ARKN-000003, ARKN-000004, ARKN-000011, ARKN-000012, ARKN-000013, ARKN-000014, ARKN-000016, ARKN-000017, ARKN-000019, ARKN-000020, ARKN-000021, ARKN-000022, ARKN-000023, ARKN-000024, ARKN-000025, ARKN-000026, ARKN-000027, ARKN-000028, ARKN-000029, ARKN-000030, ARKN-000031, ARKN-000037, ARKN-000038, ARKN-000039, ARKN-000040, ARKN-000041, ARKN-000042, ARKN-000044, ARKN-000045, ARKN-000046, ARKN-000047, ARKN-000048, ARKN-000049, ARKN-000051, ARKN-000054, ARKN-000055, ARKN-000056, ARKN-000057, ARKN-000058, ARKN-000059, ARKN-000060, ARKN-000061, ARKN-000062, ARKN-000063, ARKN-000064, ARKN-000065, ARKN-000066, ARKN-000067, ARKN-000068, ARKN-000069, ARKN-000070, ARKN-000071, ARKN-000072, ARKN-000073, ARKN-000074, ARKN-000075, ARKN-000076, ARKN-000077, ARKN-000078, ARKN-000082, ARKN-000091, ARKN-000092, ARKN-000093, ARKN-000094, ARKN-000095, ARKN-000096, ARKN-000097, ARKN-000098, ARKN-000099, ARKN-000100, ARKN-000101, ARKN-000102, ARKN-000103, ARKN-000104, ARKN-000105, ARKN-000106, ARKN-000107, ARKN-000108, ARKN-000109, ARKN-000110, ARKN-000111, ARKN-000112, ARKN-000113, ARKN-000114, ARKN-000115, ARKN-000116, ARKN-000117, ARKN-000119, ARKN-000120, ARKN-000123, ARKN-000129, ARKN-000130, ARKN-000132, ARKN-000133, ARKN-000134, ARKN-000135, ARKN-000136, ARKN-000137, ARKN-000138, ARKN-000139, ARKN-000140, ARKN-000141, ARKN-000142, ARKN-000143, ARKN-000144, ARKN-000145, ARKN-000149, ARKN-000150, ARKN-000151, ARKN-000152, ARKN-000153, ARKN-000154, ARKN-000155, ARKN-000157, ARKN-000167, ARKN-000168, arkn-000169, arkn-000170, ARKN-000171, arkn-000173, arkn-000174, arkn-000175, arkn-000176, arkn-000177, arkn-000178, arkn-000179, arkn-000180, arkn-000181, arkn-000182, arkn-000183, arkn-000184, arkn-000185, ARKN-000186, ARKN-000187, ARKN-000188, ARKN-000189, ARKN-000190, ARKN-000191, ARKN-000192, ARKN-000193, ARKN-000194, ARKN-000196, ARKN-000197, ARKN-000198, ARKN-000199, ARKN-000200, ARKN-000201, ARKN-000202, ARKN-000203, ARKN-000204, ARKN-000205, ARKN-000206, ARKN-000207, ARKN-000208, ARKN-000209, ARKN-000210, ARKN-000211, ARKN-000212, ARKN-000213, ARKN-000214, ARKN-000215, ARKN-000216, ARKN-000217, ARKN-000220, ARKN-000221, ARKN-000222, arkn-000223, arkn-000224, arkn-000226, arkn-000227, arkn-000229, arkn-000230, arkn-000231, arkn-000232, arkn-000233, ARKN-000236, arkn-000237, arkn-000238, arkn-000239, arkn-000240, arkn-000241, arkn-000242, arkn-000243, arkn-000244, ARKN-000245, arkn-000248, arkn-000249, ARKN-000250, ARKN-000251, ARKN-000252, ARKN-000253, arkn-000254, arkn-000257, arkn-000258, ARKN-000259, ARKN-000260, arkn-000261, ARKN-000262, arkn-000263, arkn-000264, arkn-000266, arkn-000269, arkn-000270, arkn-000271, arkn-000272, arkn-000273, arkn-000274, arkn-000275, arkn-000276, arkn-000277, ARKN-000278, ARKN-000279, ARKN-000280, ARKN-000281, ARKN-000282, ARKN-000283, ARKN-000284, ARKN-000285, ARKN-000286, ARKN-000287, ARKN-000288, ARKN-000289, ARKN-000290, ARKN-000293, ARKN-000294, ARKN-000295, ARKN-000296, ARKN-000297, ARKN-000300, ARKN-000301, ARKN-000302, ARKN-000303, ARKN-000304, ARKN-000305, ARKN-000306, ARKN-000307, ARKN-000308, ARKN-000309, ARKN-000310, ARKN-000311, ARKN-000312, ARKN-000313, arkn-000314, arkn-000315, arkn-000316, arkn-000317, arkn-000318, ARKN-000319, arkn-000320, arkn-000321, arkn-000322, arkn-000323, arkn-000324, arkn-000325, arkn-000327, arkn-000328, arkn-000329, arkn-000330, ARKN-000332, ARKN-000333, ARKN-000334, ARKN-000335, ARKN-000336, ARKN-000337, ARKN-000338, ARKN-000339, ARKN-000340, ARKN-000341, ARKN-000342, ARKN-000343, ARKN-000344, ARKN-000345, ARKN-000353, ARKN-000360, ARKN-000361, ARKN-000362, ARKN-000363, ARKN-000364, ARKN-000365, ARKN-000366, ARKN-000367, ARKN-000369, ARKN-000370, ARKN-000371, ARKN-000372, ARKN-000373, ARKN-000374, ARKN-000376, ARKN-000377, ARKN-000378, ARKN-000379, ARKN-000380, ARKN-000381, ARKN-000382, ARKN-000383, ARKN-000384, ARKN-000385, ARKN-000386, ARKN-000387, ARKN-000388, ARKN-000389, ARKN-000390, ARKN-000391, ARKN-000392, ARKN-000393, ARKN-000394,
Risk classification
Class II
Area (FDA)
device
Country
United Kingdom

Related recalls

Questions about this recall

Was Arkon Anesthesia Delivery System, Model 99999. Product Usage: Continuous Ventilator The Arkon Anesthesia Delivery System is designed to provide emergency oxygen, vaporized agent delivery and manual ventilation in the event of a power failure scenario. Clinicians are able to use Emergency O2 and manually ventilate the patient, providing gas and agent. However, the anesthesiologist will not have access to mechanical ventilation or ventilator monitoring. recalled?

Yes. FDA published recall Z-1460-2017 on March 22, 2017. Reported hazard: Spacelabs has received reports of the Arkon Anesthesia Delivery System, Model 99999, shutting down without warning. The power supply has failed in a way that does not trigger the normal on battery alarms.

What should I do if I have this product?

Voluntary: Firm initiated · Letter