IntelliSpace Portal DX/HX/EX AutoSPECT Pro Software Application, Philips Medical Systems, Cleveland, OH. Provides software applications used to process, analyze, and display medical images/data.
Published April 27, 2016 · Recall initiated: March 4, 2016
Summary
FDA records medical device Class II recall Z-1506-2016, published April 27, 2016 and initiated March 4, 2016. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 33 Units; Lot: M/N 881001; S/N: 85254, 82202, 85330, 82117, 7000022, 105009, 970001, 850001, 870030, 82141, 95068, 87182, 87197, 95078, 82140, 11017, 98139, 95035, 87200, 980047, 85314, 85303, 980046, 85298, 822022, 7000019, 950076, 950115, 95043, 95047, 950021, 960123 & 105006. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90® or 180® relative to one another. However, certain gamma cameras allow for other relative detector angles. Data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro and the results will likely be erroneous.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Distributor
- Worldwide Distribution -- US, including the states of CO, DE, FL, GA, IL, IN, KS, MO, NC, NV, NY, OH, OR, PA, TN & TX; and, the countries of Australia, Austria, Belgium, Canada, China, Denmark, Egypt, France, Germany, Greece, Ireland, Israel, Italy, Lebanon, Malaysia, Saudi Arabia, Slovenia, South Africa, Switzerland, Turkey, & United Kingdom.
- Lot
- M/N 881001; S/N: 85254, 82202, 85330, 82117, 7000022, 105009, 970001, 850001, 870030, 82141, 95068, 87182, 87197, 95078, 82140, 11017, 98139, 95035, 87200, 980047, 85314, 85303, 980046, 85298, 822022, 7000019, 950076, 950115, 95043, 95047, 950021, 960123 & 105006.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAPinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.September 21, 2022
- FDABrilliance 64 Model # 728231, computed tomography x-ray systemAugust 14, 2019
- FDAIngenuity Core128 Model # 728323, computed tomography x-ray systemAugust 14, 2019
- FDABrightView XCT , Model Number 882482July 31, 2019
- FDAExtended Brilliance Workspace NM Special Nuclear medicine image display and processing application suite.April 27, 2016
- FDAOssur Lil Angel Vest P3 Kodel. Pediatric sized vest is a support structure used with the halo system forApril 27, 2016
Questions about this recall
Was IntelliSpace Portal DX/HX/EX AutoSPECT Pro Software Application, Philips Medical Systems, Cleveland, OH. Provides software applications used to process, analyze, and display medical images/data. recalled?
Yes. FDA published recall Z-1506-2016 on April 27, 2016. Reported hazard: The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90® or 180® relative to one another. However, certain gamma cameras allow for other relative detector angles. Data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro and the results will likely be erroneous.
What should I do if I have this product?
Voluntary: Firm initiated · Letter