FDAFDA · Food and drugs·Recall number Z-1505-2016
Extended Brilliance Workspace NM Special Nuclear medicine image display and processing application suite. AutoSPECT Pro Software Application. Philips Medical Systems, Cleveland, OH. Provides software applications used to process, analyze, and display medical images/data.
Published April 27, 2016 · Recall initiated: March 4, 2016
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90¿ or 180¿ relative to one another. However, certain gamma cameras allow for other relative detector angles. Data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro and the results will likely be erroneous.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Philips Medical Systems (Cleveland) Inc
- Distributor
- Worldwide Distribution -- US, including the states of CO, DE, FL, GA, IL, IN, KS, MO, NC, NV, NY, OH, OR, PA, TN & TX; and, the countries of Australia, Austria, Belgium, Canada, China, Denmark, Egypt, France, Germany, Greece, Ireland, Israel, Italy, Lebanon, Malaysia, Saudi Arabia, Slovenia, South Africa, Switzerland, Turkey, & United Kingdom.
- Lot
- M/N 882488; S/N: 66565, 66586, 66589, 66670, 66681, 66672, 66644, 66888, 66374, 66371, 66305, 66824, 66261, 66356, 66059, 66056, 67027, 66248, 66590, 66240, 66097, 66268, 66555, 66500, 66754, 66469, 66411, 66252, 66263, 66307, 66311, 66310, 66306, 66259, 67054, 67000 & 67017.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States