Extended Brilliance Workspace NM Special Nuclear medicine image display and processing application suite. AutoSPECT Pro Software Application. Philips Medical Systems, Cleveland, OH. Provides software applications used to process, analyze, and display medical images/data.

Published April 27, 2016 · Recall initiated: March 4, 2016

Summary

FDA records medical device Class II recall Z-1505-2016, published April 27, 2016 and initiated March 4, 2016. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 37 Units; Lot: M/N 882488; S/N: 66565, 66586, 66589, 66670, 66681, 66672, 66644, 66888, 66374, 66371, 66305, 66824, 66261, 66356, 66059, 66056, 67027, 66248, 66590, 66240, 66097, 66268, 66555, 66500, 66754, 66469, 66411, 66252, 66263, 66307, 66311, 66310, 66306, 66259, 67054, 67000 & 67017. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90® or 180® relative to one another. However, certain gamma cameras allow for other relative detector angles. Data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro and the results will likely be erroneous.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips Medical Systems (Cleveland) Inc
Distributor
Worldwide Distribution -- US, including the states of CO, DE, FL, GA, IL, IN, KS, MO, NC, NV, NY, OH, OR, PA, TN & TX; and, the countries of Australia, Austria, Belgium, Canada, China, Denmark, Egypt, France, Germany, Greece, Ireland, Israel, Italy, Lebanon, Malaysia, Saudi Arabia, Slovenia, South Africa, Switzerland, Turkey, & United Kingdom.
Lot
M/N 882488; S/N: 66565, 66586, 66589, 66670, 66681, 66672, 66644, 66888, 66374, 66371, 66305, 66824, 66261, 66356, 66059, 66056, 67027, 66248, 66590, 66240, 66097, 66268, 66555, 66500, 66754, 66469, 66411, 66252, 66263, 66307, 66311, 66310, 66306, 66259, 67054, 67000 & 67017.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Extended Brilliance Workspace NM Special Nuclear medicine image display and processing application suite. AutoSPECT Pro Software Application. Philips Medical Systems, Cleveland, OH. Provides software applications used to process, analyze, and display medical images/data. recalled?

Yes. FDA published recall Z-1505-2016 on April 27, 2016. Reported hazard: The AutoSPECT Pro application was only designed to reconstruct cardiac SPECT data obtained with detectors positioned at 90® or 180® relative to one another. However, certain gamma cameras allow for other relative detector angles. Data acquired at these other angles will not be correctly reconstructed by AutoSPECT Pro and the results will likely be erroneous.

What should I do if I have this product?

Voluntary: Firm initiated · Letter