Nuclear gamma cameras (ADAC Vertex Plus, ADAC Solus, ADAC Vertex V60 & ADAC Vertex Classic) Philips Medical System, Cleveland, OH 44143. Used to produce images depicting the anatomical distribution of single photon and positron emitting radioisotopes with the human body for interpretation by medical personnel.

Published May 7, 2014 · Recall initiated: March 10, 2014

Summary

FDA records medical device Class II recall Z-1510-2014, published May 7, 2014 and initiated March 10, 2014. Recalling firm: Philips Medical Systems (Cleveland) Inc; Status: Terminated; Reported quantity: 1,043 cameras; Lot: ADAC Vertex Plus & ADAC Solus System Code: 88297, Serial #'s: 11, 1006, 1008, 1027, 1092, 1097, 2035, 2038, 2047, 3059, 3060, 3064, 4007, 4078, 6052, 6261, 8002, 10252, 10630, 11090, 85247, 513519, 513520, 940384, 9205345, 9408136, 9503084, 9506016, 9511123, 9512009, 9512018, 9601012, 9601015, 9601016, 9601024, 9601036, 9602161, 9602254, 9603336, 9604016, 9604021, 9605024, 9605026, 9605037, 9606211, 9607015, 9607024, 9608031, 9608046, 9608049, 9608052, 9609063, 9609081, 9610058, 9611035, 9611077, 9611082, 9612072, 9702095, 9702150, 9703154, 9703165, 9704044, 9704178, 9704183, 9705004, 9705186, 9706005, 9706182, 9706189, 9707002, 9707047, 9707050, 9707054, 9708062, 9708065, 9708076, 9708078, 9708084, 9708088, 9708227, 9709022, 9709061, 9710121, 9710127, 9710244, 9711025, 9711034, 9711036, 9711074, 9711271, 9711291, 9801047, 9801227, 9801232, 9801297, 9802050, 9802255, 9802260, 9802300, 9802306, 9802397, 9803003, 9803014, 9803067, 9803083, 9803269, 9803272, 9803288, 9803290, 9803293, 9803295, 9803303, 9803312, 9803316, 9804069, 9804330, 9804333, 9804342, 9805344, 9805350, 9805376, 9806014, 9806363, 9806377, 9806388, 9807391, 9807421, 9808036, 9808090, 9808108, 9808451, 9808452, 9808457, 9808480, 9808829, 9809067, 9809113, 9809122, 9809125, 9809498, 9809687, 9810075, 9810136, 9810551, 9810557, 9811020, 9811093, 9811569, 9811737, 9812582, 9901162, 9901597, 9901605, 9902150, 9902622, 9902630, 9903632, 9903637, 9903641, 9904089, 9905003, 9905012, 9905659, 9905667, 9906670, 9907016, 9907019, 9909011, 9909161, 9910015, 9910017, 9910031, 9911458, 9911698, 9911702, 9911711, 9912181, 9912463, 98066002, 802080322, 0002032, 0007001M, 0008009B, 0011002B, 0101002C, 0102019V, 0103020V, 01069VM, 01070VM, 0111001V, 01295VM, 01296K, 01864VM, 03026VM, 0939VM, 100098-01N, 130234-01NVERTEX, 5934V, 6090VM, 7373VM, 7376VM, 7386VM, 7560VM, 7561VM, 9054VM/V, 9056VM/V, 9130VM, 9806373A, 9810428ADAC#7973VM, 9894K, AD1261, AMT00848VM, B0006129, B0007273, B0008001, B0008003, B0008012, B0008021, B0008030, B001200212, B0102114, B0104125, B0104128, B0105138, B0106169, B01070191, B01070192, B0107179, B01080199, B01080204, B01080207, B01090227, B01090228, B01100231, B01100232, B01100233, B01100236, B01100237, B01100238, B01100241, B01100242, B01110251, B01110252, B02050071, B02050262, B02050263, B02050265, B02060167, B02060270, B02060279, B0206027S, B02070289, B02070290, B02070291, B02070297, B02070300, B0207296, B02080318, B0208312, B02090327, B02090334, B02100345, B02110348, B02120352, B03010355, B03010359, B03020048, B03080380, C0006050, C0006051, C0007060, C0008070, C0009075, C0101099, C0102103, C0102108, C0104118, C0107145, C01080152, C01080153, C01080154, C02100218, C02110220, C02120223, C03010227, C03050230, C03060193, CNF, M@10458, M@10482, M@10547, M@105561, M@10573, M@10582, M@10589, M@10591, M@10676, M@10772, M@10837, M@10839, M@10848, M@10850, M@10871, M@10878, M@10882, M@10927, M@10937, M@10967, M@10979, M@11015, M@11020, M@11036, M@11103, M@11109, M@11140, M@11178, M@11184, M@11381, M@11398, M@11423, M@11498, M@11525, M@11571, M@11590, M@11609, M@11644, M@11881, M@11939, M@11949, M@11957, M@11972, M@12060, M@12219, M@12412, M@12591, M@12711, M@12787, M@13010, M@13296, M@13466, M@13498, M@13543, M@13549, M@13663, M@13665, M@13688, M@13691, M@13748, M@13813, M@14062, M@22752, M@28051, M@4750019856, M@5828J, M@5906N, M@5919J, M@5936N, M@5956J, M@6002J, M@6012J, M@6075VM, M@6076K, M@6077K, M@6108VA, M@6123VA, M@6143VA, M@6160VM, M@6183VA, M@6244VA, M@6258J, M@6323VM, M@6324J, M@6330VA, M@6341VA, M@6344VM, M@6345VA, M@6346VM, M@6348VM, M@6349VM, M@6365VM, M@6366K, M@6368J, M@6370VA, M@6371VM, M@6375J, M@6376J, M@6460VM, M@6470VA, M@6481VA, M@6496VM, M@6528VA, M@6541VA, M@6542VM, M@6566VM, M@6573VM, M@6596VM, M@6600VA, M@6638VM, M@6639J, M@6675VA, M@6716VM, M@6717VM, M@6762VM, M@6769K, M@6771VM, M@6773VM, M@6776VM, M@6797K, M@6805VM, M@6823VM, M@6840K, M@6841VM, M@6867VM, M@6877VM, M@6886VM, M@6891VA, M@6912K, M@6924VM, M@6931VM, M@6932VM, M@6943VM, M@6. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Slippage of the radius drive belt in the relative 180 degree orientation of the superior positioned detector head will allow the detector head to drift down and potentially come in direct contact with the patient.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Philips Medical Systems (Cleveland) Inc
Distributor
Worldwide Distribution -- USA, and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Denmark, Egypt, France, Germany, Hong Kong, Ireland, Italy, Korea, Lebanon, Malaysia, Netherlands, Norway, Pakistan, Portugal, Russia, Singapore, Spain, Sweden, Thailand, & United Kingdom.
Lot
ADAC Vertex Plus & ADAC Solus System Code: 88297, Serial #'s: 11, 1006, 1008, 1027, 1092, 1097, 2035, 2038, 2047, 3059, 3060, 3064, 4007, 4078, 6052, 6261, 8002, 10252, 10630, 11090, 85247, 513519, 513520, 940384, 9205345, 9408136, 9503084, 9506016, 9511123, 9512009, 9512018, 9601012, 9601015, 9601016, 9601024, 9601036, 9602161, 9602254, 9603336, 9604016, 9604021, 9605024, 9605026, 9605037, 9606211, 9607015, 9607024, 9608031, 9608046, 9608049, 9608052, 9609063, 9609081, 9610058, 9611035, 9611077, 9611082, 9612072, 9702095, 9702150, 9703154, 9703165, 9704044, 9704178, 9704183, 9705004, 9705186, 9706005, 9706182, 9706189, 9707002, 9707047, 9707050, 9707054, 9708062, 9708065, 9708076, 9708078, 9708084, 9708088, 9708227, 9709022, 9709061, 9710121, 9710127, 9710244, 9711025, 9711034, 9711036, 9711074, 9711271, 9711291, 9801047, 9801227, 9801232, 9801297, 9802050, 9802255, 9802260, 9802300, 9802306, 9802397, 9803003, 9803014, 9803067, 9803083, 9803269, 9803272, 9803288, 9803290, 9803293, 9803295, 9803303, 9803312, 9803316, 9804069, 9804330, 9804333, 9804342, 9805344, 9805350, 9805376, 9806014, 9806363, 9806377, 9806388, 9807391, 9807421, 9808036, 9808090, 9808108, 9808451, 9808452, 9808457, 9808480, 9808829, 9809067, 9809113, 9809122, 9809125, 9809498, 9809687, 9810075, 9810136, 9810551, 9810557, 9811020, 9811093, 9811569, 9811737, 9812582, 9901162, 9901597, 9901605, 9902150, 9902622, 9902630, 9903632, 9903637, 9903641, 9904089, 9905003, 9905012, 9905659, 9905667, 9906670, 9907016, 9907019, 9909011, 9909161, 9910015, 9910017, 9910031, 9911458, 9911698, 9911702, 9911711, 9912181, 9912463, 98066002, 802080322, 0002032, 0007001M, 0008009B, 0011002B, 0101002C, 0102019V, 0103020V, 01069VM, 01070VM, 0111001V, 01295VM, 01296K, 01864VM, 03026VM, 0939VM, 100098-01N, 130234-01NVERTEX, 5934V, 6090VM, 7373VM, 7376VM, 7386VM, 7560VM, 7561VM, 9054VM/V, 9056VM/V, 9130VM, 9806373A, 9810428ADAC#7973VM, 9894K, AD1261, AMT00848VM, B0006129, B0007273, B0008001, B0008003, B0008012, B0008021, B0008030, B001200212, B0102114, B0104125, B0104128, B0105138, B0106169, B01070191, B01070192, B0107179, B01080199, B01080204, B01080207, B01090227, B01090228, B01100231, B01100232, B01100233, B01100236, B01100237, B01100238, B01100241, B01100242, B01110251, B01110252, B02050071, B02050262, B02050263, B02050265, B02060167, B02060270, B02060279, B0206027S, B02070289, B02070290, B02070291, B02070297, B02070300, B0207296, B02080318, B0208312, B02090327, B02090334, B02100345, B02110348, B02120352, B03010355, B03010359, B03020048, B03080380, C0006050, C0006051, C0007060, C0008070, C0009075, C0101099, C0102103, C0102108, C0104118, C0107145, C01080152, C01080153, C01080154, C02100218, C02110220, C02120223, C03010227, C03050230, C03060193, CNF, M@10458, M@10482, M@10547, M@105561, M@10573, M@10582, M@10589, M@10591, M@10676, M@10772, M@10837, M@10839, M@10848, M@10850, M@10871, M@10878, M@10882, M@10927, M@10937, M@10967, M@10979, M@11015, M@11020, M@11036, M@11103, M@11109, M@11140, M@11178, M@11184, M@11381, M@11398, M@11423, M@11498, M@11525, M@11571, M@11590, M@11609, M@11644, M@11881, M@11939, M@11949, M@11957, M@11972, M@12060, M@12219, M@12412, M@12591, M@12711, M@12787, M@13010, M@13296, M@13466, M@13498, M@13543, M@13549, M@13663, M@13665, M@13688, M@13691, M@13748, M@13813, M@14062, M@22752, M@28051, M@4750019856, M@5828J, M@5906N, M@5919J, M@5936N, M@5956J, M@6002J, M@6012J, M@6075VM, M@6076K, M@6077K, M@6108VA, M@6123VA, M@6143VA, M@6160VM, M@6183VA, M@6244VA, M@6258J, M@6323VM, M@6324J, M@6330VA, M@6341VA, M@6344VM, M@6345VA, M@6346VM, M@6348VM, M@6349VM, M@6365VM, M@6366K, M@6368J, M@6370VA, M@6371VM, M@6375J, M@6376J, M@6460VM, M@6470VA, M@6481VA, M@6496VM, M@6528VA, M@6541VA, M@6542VM, M@6566VM, M@6573VM, M@6596VM, M@6600VA, M@6638VM, M@6639J, M@6675VA, M@6716VM, M@6717VM, M@6762VM, M@6769K, M@6771VM, M@6773VM, M@6776VM, M@6797K, M@6805VM, M@6823VM, M@6840K, M@6841VM, M@6867VM, M@6877VM, M@6886VM, M@6891VA, M@6912K, M@6924VM, M@6931VM, M@6932VM, M@6943VM, M@6
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Nuclear gamma cameras (ADAC Vertex Plus, ADAC Solus, ADAC Vertex V60 & ADAC Vertex Classic) Philips Medical System, Cleveland, OH 44143. Used to produce images depicting the anatomical distribution of single photon and positron emitting radioisotopes with the human body for interpretation by medical personnel. recalled?

Yes. FDA published recall Z-1510-2014 on May 7, 2014. Reported hazard: Slippage of the radius drive belt in the relative 180 degree orientation of the superior positioned detector head will allow the detector head to drift down and potentially come in direct contact with the patient.

What should I do if I have this product?

Voluntary: Firm initiated · Letter