Hospira Acclaim Encore Infusion Pump Designed to deliver parenteral infusions including whole blood or red blood cell components and enteral fluids using a wide variety of standard administration sets and fluid containers.
Published May 7, 2014 · Recall initiated: October 31, 2013
Summary
FDA records medical device Class I recall Z-1483-2014, published May 7, 2014 and initiated October 31, 2013. Recalling firm: Hospira Inc.; Status: Terminated; Reported quantity: 13,218 infusion pumps total; Lot: List Number: 12237; Serial Numbers: 12670196, 12670182, 12670203, 12670281, 12671061, 12671058, 12671175, 12718477, 12672450, 12672413, 12716099, 12716071, 12671563, 12672061, 12672079, 12672156, 12718948, 12717835, 12716259, 12716268, 12671580, 12671554, 12716244, 12716293, 12671647 , 12672014 , 12672021 , 12672064 , 12672074 , 12672137 , 12672138 , 12672143 , 12672173 , 12672179 , 12672185 , 12672186 , 12672161 , 12672217 , 12716140 , 12716324 , 12716396 , 12716489 , 12716316 , 12671666 , 12672065 , 12672175 , 12716251 , 12716550 , 12716398 , 12716963 , 12716451 , 12716345 , 12716519 , 12716693 , 12716477 , 12716610 , 12716638 , 12716344 , 12716538 , 12716515 , 12716315 , 12716720 , 12716623 , 12716787 , 12716822 , 12717186 , 12716738 , 12717343 , 12716798 , 12716812 , 12716532 , 12717123 , 12716849 , 12717281 , 12717049 , 12717160 , 12717359 , 12717028 , 12717298 , 12673508 , 12673609 , 12673114 , 12673524 , 12673629 , 12673529 , 12673563 , 12673597 , 12673610 , 12673700 , 12673741 , 12673584 , 12673647 , 12673679 , 12673711 , 12672796 , 12672848 , 12672740 , 12672801 , 12673033 , 12673099 , 12671687 , 12673054 , 12673127 , 12673035 , 12673144 , 12672775 , 12673082 , 12673131 , 12673638 , 12672747 , 12672784 , 12672789 , 12673069 , 12673083 , 12673159 , 12673180 , 12673350 , 12673443 , 12718788 , 12716217 , 12672791 , 12672794 , 12672766 , 12672925 , 12672990 , 12673212 , 12673165 , 12672749 , 12672777 , 12672882 , 12672886 , 12671505 , 12673916 , 12717412 , 12673887 , 12673899 , 12673919 , 12673922 , 12673915 , 12673926 , 12673947 , 12674013 , 12674024 , 12672845 , 12673050 , 12674010 , 12673971 , 12716417 , 12717963 , 12718476 , 12672244 , 12716470 , 12671680 , 12717834 , 12718964 , 12718713 , 12673944 , 12717693 , 12671566 , 12672008 , 12674347 , 12674349 , 12674392 , 12674416 , 12671926 , 12674363 , 12674380 , 12674428 , 12674475 , 12674534 , 12674362 , 12674407 , 12674430 , 12674443 , 12716672 , 12671859 , 12717799 , 12717966 , 12674144 , 12674098 , 12674119 , 12674086 , 12674090 , 12674110 , 12674117 , 12674137 , 12674244 , 12674235 , 12674170 , 12674263 , 12674163 , 12674269 , 12674289 , 12674234 , 12674344 , 12674783 , 12674851 , 12674860 , 12674868 , 12674886 , 12674925 , 12674942 , 12674955 , 12675038 , 12716560 , 12674328 , 12674885 , 12674785 , 12674787 , 12674794 , 12674799 , 12674812 , 12674813 , 12674814 , 12674821 , 12674830 , 12674833 , 12674852 , 12674855 , 12674895 , 12674899 , 12674911 , 12674913 , 12674928 , 12674961 , 12674975 , 12674976 , 12674981 , 12674982 , 12674988 , 12674989 , 12674992 , 12674993 , 12674998 , 12675001 , 12675005 , 12675007 , 12675008 , 12675010 , 12675015 , 12675016 , 12675017 , 12675018 , 12675023 , 12675025 , 12675026 , 12675030 , 12675032 , 12675035 , 12675036 , 12675044 , 12675045 , 12675046 , 12675047 , 12675054 , 12675055 , 12675056 , 12675057 , 12675060 , 12675061 , 12675063 , 12675064 , 12675065 , 12675068 , 12675070 , 12675072 , 12675073 , 12675074 , 12675075 , 12675076 , 12675077 , 12675079 , 12675080 , 12675083 , 12675084 , 12675086 , 12675087 , 12675088 , 12675089 , 12675093 , 12675095 , 12675096 , 12675097 , 12675098 , 12675099 , 12675101 , 12675103 , 12675104 , 12675106 , 12675107 , 12675108 , 12675109 , 12675112 , 12675113 , 12675114 , 12675115 , 12675119 , 12675121 , 12675122 , 12675124 , 12675125 , 12675126 , 12675127 , 12675129 , 12675132 , 12675133 , 12675134 , 12675135 , 12675136 , 12675141 , 12675144 , 12675146 , 12675147 , 12675148 , 12674792 , 12674801 , 12674844 , 12674859 , 12674872 , 12674874 , 12674889 , 12674931 , 12674959 , 12674968 , 12675168 , 12674983 , 12673495 , 12674781 , 12674797 , 12674906 , 12674920 , 12674960 , 12. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Broken door assemblies on the Hospira/Abbott Acclaim Encore infusion pumps. If the door assembly breaks, it may prevent the door from closing properly and unrestricted flow may occur. If the door cannot be closed, the pump cannot be used which can result in a delay in therapy.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Hospira Inc.
- Distributor
- Nationwide Distribution -- AK, AL, AR, AZ, CA, CO, CR, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY.
- Lot
- List Number: 12237; Serial Numbers: 12670196, 12670182, 12670203, 12670281, 12671061, 12671058, 12671175, 12718477, 12672450, 12672413, 12716099, 12716071, 12671563, 12672061, 12672079, 12672156, 12718948, 12717835, 12716259, 12716268, 12671580, 12671554, 12716244, 12716293, 12671647 , 12672014 , 12672021 , 12672064 , 12672074 , 12672137 , 12672138 , 12672143 , 12672173 , 12672179 , 12672185 , 12672186 , 12672161 , 12672217 , 12716140 , 12716324 , 12716396 , 12716489 , 12716316 , 12671666 , 12672065 , 12672175 , 12716251 , 12716550 , 12716398 , 12716963 , 12716451 , 12716345 , 12716519 , 12716693 , 12716477 , 12716610 , 12716638 , 12716344 , 12716538 , 12716515 , 12716315 , 12716720 , 12716623 , 12716787 , 12716822 , 12717186 , 12716738 , 12717343 , 12716798 , 12716812 , 12716532 , 12717123 , 12716849 , 12717281 , 12717049 , 12717160 , 12717359 , 12717028 , 12717298 , 12673508 , 12673609 , 12673114 , 12673524 , 12673629 , 12673529 , 12673563 , 12673597 , 12673610 , 12673700 , 12673741 , 12673584 , 12673647 , 12673679 , 12673711 , 12672796 , 12672848 , 12672740 , 12672801 , 12673033 , 12673099 , 12671687 , 12673054 , 12673127 , 12673035 , 12673144 , 12672775 , 12673082 , 12673131 , 12673638 , 12672747 , 12672784 , 12672789 , 12673069 , 12673083 , 12673159 , 12673180 , 12673350 , 12673443 , 12718788 , 12716217 , 12672791 , 12672794 , 12672766 , 12672925 , 12672990 , 12673212 , 12673165 , 12672749 , 12672777 , 12672882 , 12672886 , 12671505 , 12673916 , 12717412 , 12673887 , 12673899 , 12673919 , 12673922 , 12673915 , 12673926 , 12673947 , 12674013 , 12674024 , 12672845 , 12673050 , 12674010 , 12673971 , 12716417 , 12717963 , 12718476 , 12672244 , 12716470 , 12671680 , 12717834 , 12718964 , 12718713 , 12673944 , 12717693 , 12671566 , 12672008 , 12674347 , 12674349 , 12674392 , 12674416 , 12671926 , 12674363 , 12674380 , 12674428 , 12674475 , 12674534 , 12674362 , 12674407 , 12674430 , 12674443 , 12716672 , 12671859 , 12717799 , 12717966 , 12674144 , 12674098 , 12674119 , 12674086 , 12674090 , 12674110 , 12674117 , 12674137 , 12674244 , 12674235 , 12674170 , 12674263 , 12674163 , 12674269 , 12674289 , 12674234 , 12674344 , 12674783 , 12674851 , 12674860 , 12674868 , 12674886 , 12674925 , 12674942 , 12674955 , 12675038 , 12716560 , 12674328 , 12674885 , 12674785 , 12674787 , 12674794 , 12674799 , 12674812 , 12674813 , 12674814 , 12674821 , 12674830 , 12674833 , 12674852 , 12674855 , 12674895 , 12674899 , 12674911 , 12674913 , 12674928 , 12674961 , 12674975 , 12674976 , 12674981 , 12674982 , 12674988 , 12674989 , 12674992 , 12674993 , 12674998 , 12675001 , 12675005 , 12675007 , 12675008 , 12675010 , 12675015 , 12675016 , 12675017 , 12675018 , 12675023 , 12675025 , 12675026 , 12675030 , 12675032 , 12675035 , 12675036 , 12675044 , 12675045 , 12675046 , 12675047 , 12675054 , 12675055 , 12675056 , 12675057 , 12675060 , 12675061 , 12675063 , 12675064 , 12675065 , 12675068 , 12675070 , 12675072 , 12675073 , 12675074 , 12675075 , 12675076 , 12675077 , 12675079 , 12675080 , 12675083 , 12675084 , 12675086 , 12675087 , 12675088 , 12675089 , 12675093 , 12675095 , 12675096 , 12675097 , 12675098 , 12675099 , 12675101 , 12675103 , 12675104 , 12675106 , 12675107 , 12675108 , 12675109 , 12675112 , 12675113 , 12675114 , 12675115 , 12675119 , 12675121 , 12675122 , 12675124 , 12675125 , 12675126 , 12675127 , 12675129 , 12675132 , 12675133 , 12675134 , 12675135 , 12675136 , 12675141 , 12675144 , 12675146 , 12675147 , 12675148 , 12674792 , 12674801 , 12674844 , 12674859 , 12674872 , 12674874 , 12674889 , 12674931 , 12674959 , 12674968 , 12675168 , 12674983 , 12673495 , 12674781 , 12674797 , 12674906 , 12674920 , 12674960 , 12
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAInfant 25% DEXTROSE Injection, USP, 2.5 g (250 mg/mL), 10 mL Unit of Use Single-dose Syringe per carton, RxMay 31, 2017
- FDAMinibore Extension Set, 59 Inch with Spin Lock Collar, Non-DEHP, Latex-Free Intended use: For theJanuary 18, 2017
- FDAFentanyl Citrate Injection, USP, 100 mcg Fentanyl/ 2mL, Single Dose, 50 mcg/mL, Rx only, Hospira, Inc., LakeNovember 30, 2016
- FDASymbiq Two Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids. List numberNovember 16, 2016
- FDAAbbott Acclaim Infusion Pump Designed to deliver parenteral infusions including whole blood or red blood cellMay 7, 2014
- FDANuclear gamma cameras (ADAC Vertex Plus, ADAC Solus, ADAC Vertex V60 & ADAC Vertex Classic) Philips MedicalMay 7, 2014
Questions about this recall
Was Hospira Acclaim Encore Infusion Pump Designed to deliver parenteral infusions including whole blood or red blood cell components and enteral fluids using a wide variety of standard administration sets and fluid containers. recalled?
Yes. FDA published recall Z-1483-2014 on May 7, 2014. Reported hazard: Broken door assemblies on the Hospira/Abbott Acclaim Encore infusion pumps. If the door assembly breaks, it may prevent the door from closing properly and unrestricted flow may occur. If the door cannot be closed, the pump cannot be used which can result in a delay in therapy.
What should I do if I have this product?
Voluntary: Firm initiated · Letter