Echelon, Echelon Oval and Oasis MR imaging device systems; one unit per box. The Echelon, Echelon Oval and Oasis MR systems are imaging devices, and are intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR systems produce transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities.
Published June 26, 2013 · Recall initiated: January 31, 2013
Summary
FDA records medical device Class II recall Z-1559-2013, published June 26, 2013 and initiated January 31, 2013. Recalling firm: Hitachi Medical Systems America Inc; Status: Terminated; Reported quantity: 45 system units; Lot: The following serial numbers are associated with this recall: Echelon - V061, V062, and V065 Oasis - M009, M017, M067, M071, M084, M106-M108, M110-M141 inclusive, and M952 Echelon Oval - Y001-Y005 inclusive. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Hitachi discovered a software error that can occur when simultaneously scanning a patient while processing images previously acquired from a different patient. The software error can occur randomly. The error causes an image in the processed dataset to have the same unique identifier (UID) as in the scanning dataset. The processed image is stored in the place of the original image, which is lost.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Hitachi Medical Systems America Inc
- Distributor
- Worldwide Distribution - US Nationwide including the states of AZ, CA, CO, DE, FL, IL, IN, KS, KY, MD, NC, ND, NY, OH, OR, PA, TX, VA, WA, and WI. One (1) system unit was also distributed to Mexico.
- Lot
- The following serial numbers are associated with this recall: Echelon - V061, V062, and V065 Oasis - M009, M017, M067, M071, M084, M106-M108, M110-M141 inclusive, and M952 Echelon Oval - Y001-Y005 inclusive
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHitachi Echelon MRI SystemMarch 29, 2017
- FDAHitachi Oasis MRI SystemMarch 29, 2017
- FDAHitachi Oasis MRI System - C-Spine CoilFebruary 1, 2017
- FDAHitachi Echelon Oval MRI system is a diagnostic imaging device (one unit per package) and is intended toSeptember 10, 2014
- FDAGlideScope Video Laryngoscope (GVL) 3, 4, 5 and GlideScope (AVL) 2, 3, 4 and 5 reusable blades Product Usage:June 26, 2013
- FDATecan Freedom EVO with Touch Tools Suite Version 3.0 The intended use of the Touch Tools Suite is to runJune 26, 2013
Questions about this recall
Was Echelon, Echelon Oval and Oasis MR imaging device systems; one unit per box. The Echelon, Echelon Oval and Oasis MR systems are imaging devices, and are intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR systems produce transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities. recalled?
Yes. FDA published recall Z-1559-2013 on June 26, 2013. Reported hazard: Hitachi discovered a software error that can occur when simultaneously scanning a patient while processing images previously acquired from a different patient. The software error can occur randomly. The error causes an image in the processed dataset to have the same unique identifier (UID) as in the scanning dataset. The processed image is stored in the place of the original image, which is lost.
What should I do if I have this product?
Voluntary: Firm initiated · Letter