Echelon, Echelon Oval and Oasis MR imaging device systems; one unit per box. The Echelon, Echelon Oval and Oasis MR systems are imaging devices, and are intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. The MR systems produce transverse, coronal, sagittal, oblique, and curved cross-sectional images that display the internal structure of the head, body, or extremities.

Published June 26, 2013 · Recall initiated: January 31, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Hitachi discovered a software error that can occur when simultaneously scanning a patient while processing images previously acquired from a different patient. The software error can occur randomly. The error causes an image in the processed dataset to have the same unique identifier (UID) as in the scanning dataset. The processed image is stored in the place of the original image, which is lost.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Hitachi Medical Systems America Inc
Distributor
Worldwide Distribution - US Nationwide including the states of AZ, CA, CO, DE, FL, IL, IN, KS, KY, MD, NC, ND, NY, OH, OR, PA, TX, VA, WA, and WI. One (1) system unit was also distributed to Mexico.
Lot
The following serial numbers are associated with this recall: Echelon - V061, V062, and V065 Oasis - M009, M017, M067, M071, M084, M106-M108, M110-M141 inclusive, and M952 Echelon Oval - Y001-Y005 inclusive
Risk classification
Class II
Area (FDA)
device
Country
United States