Hitachi Echelon Oval MRI system is a diagnostic imaging device (one unit per package) and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation.
Published September 10, 2014 · Recall initiated: August 18, 2014
Summary
FDA records medical device Class II recall Z-2564-2014, published September 10, 2014 and initiated August 18, 2014. Recalling firm: Hitachi Medical Systems America Inc; Status: Terminated; Reported quantity: 39 systems; Lot: Product Codes: Y001, Y002, Y003, Y004, Y005, Y006, Y007, Y008, Y009, Y010, Y011, Y012, Y014, Y015, Y016, Y101, Y102, Y103, Y105, and Y951. To be updated as firm submits information. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Gradient Coil was found to have a failure mode that allowed it to overheat and become a burn hazard.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Hitachi Medical Systems America Inc
- Distributor
- Worldwide Distribution: US (nationwide) including states of: CA, DE, FL, ID, KS, KY, MD, MS, NY, OH, and WY; countries of: Japan, Brazil, France, Germany, and Kyrgyzstan.
- Lot
- Product Codes: Y001, Y002, Y003, Y004, Y005, Y006, Y007, Y008, Y009, Y010, Y011, Y012, Y014, Y015, Y016, Y101, Y102, Y103, Y105, and Y951. To be updated as firm submits information.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAHitachi Echelon MRI SystemMarch 29, 2017
- FDAHitachi Oasis MRI SystemMarch 29, 2017
- FDAHitachi Oasis MRI System - C-Spine CoilFebruary 1, 2017
- FDARoche cobas IT 1000 Product Usage: Intended Use: An electronic device intended to store, retrieve, and processSeptember 10, 2014
- FDAMedsolutions4U Multi-Function Electrodes to be used with PhysioControl Lifepak Defibrillators, models 9, 10C,September 10, 2014
- FDAEchelon, Echelon Oval and Oasis MR imaging device systems; one unit per box. The Echelon, Echelon Oval andJune 26, 2013
Questions about this recall
Was Hitachi Echelon Oval MRI system is a diagnostic imaging device (one unit per package) and is intended to provide the physician with physiological and clinical information, obtained non-invasively and without the use of ionizing radiation. recalled?
Yes. FDA published recall Z-2564-2014 on September 10, 2014. Reported hazard: The Gradient Coil was found to have a failure mode that allowed it to overheat and become a burn hazard.
What should I do if I have this product?
Voluntary: Firm initiated · Letter