Panorama Patient Monitoring Network Mindray DS USA The Panorama Patient Monitoring can view recal time, store, print, graph and trend patient clinical and demographic data. The Panoram Patient Monitoring Network can sent independent alarm limits for data send by the bedside monitor.
Published May 4, 2016 · Recall initiated: March 9, 2016
Summary
FDA records medical device Class II recall Z-1575-2016, published May 4, 2016 and initiated March 9, 2016. Recalling firm: Mindray DS USA, Inc. dba Mindray North America; Status: Terminated; Reported quantity: 1447 units US, 45 units OUS; Lot: Panorama Central Station's CPU Unit, Tower (p/n 0998-00-0709-01) Panorama Central Station's CPU Unit, 2U case (p/n 0998-00-0708-01) Panorama Centrall Station's eGateway Tower (p/n 0998-00-0709-03) Panorama Central Station's eGateway 2U case (p/n 0998-00-0708-03). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Panorama Central Station including the work station View Station, View Station and the eGateway will revert to the year 2000 when Daylight Savings Time (DST) occurs on March 13, 2016, or under various conditions.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Mindray DS USA, Inc. dba Mindray North America
- Distributor
- Nationwide Distribution
- Lot
- Panorama Central Station's CPU Unit, Tower (p/n 0998-00-0709-01) Panorama Central Station's CPU Unit, 2U case (p/n 0998-00-0708-01) Panorama Centrall Station's eGateway Tower (p/n 0998-00-0709-03) Panorama Central Station's eGateway 2U case (p/n 0998-00-0708-03)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASevoflurane (SEVO) V90 Electronic Vaporizer with the A9 Anesthesia System. Labeled as the following: 1. SevDecember 3, 2025
- FDADesflurane E_Vap.(Saf-T-Fill) used with the A9 Anesthesia System-Used to provide the vapor of an anestheticSeptember 11, 2024
- FDAIsoflurane E_Vap.(Safety Filling adapter): used with the A9 Anesthesia System-Used to provide the vapor of anSeptember 11, 2024
- FDASevoflurane E_Vap.(Safety Filling adapter) used with the A9 Anesthesia System-Used to provide the vapor of anSeptember 11, 2024
- FDABD Trucount Absolute Counting Tubes: Model number: 340334 IVD; BD Trucount tubes are used for determiningMay 4, 2016
- FDAHologic InPlex CF Molecular Test (IVD); Catalog Number: 95-432.May 4, 2016
Questions about this recall
Was Panorama Patient Monitoring Network Mindray DS USA The Panorama Patient Monitoring can view recal time, store, print, graph and trend patient clinical and demographic data. The Panoram Patient Monitoring Network can sent independent alarm limits for data send by the bedside monitor. recalled?
Yes. FDA published recall Z-1575-2016 on May 4, 2016. Reported hazard: Panorama Central Station including the work station View Station, View Station and the eGateway will revert to the year 2000 when Daylight Savings Time (DST) occurs on March 13, 2016, or under various conditions.
What should I do if I have this product?
Voluntary: Firm initiated · Letter