FDAFDA · Food and drugs·Recall number Z-1583-2014

DRAPE 9732722 TUBE STERILE O-ARM 20PK

Published November 26, 2014 · Recall initiated: April 24, 2014

Summary

FDA records medical device Class II recall Z-1583-2014, published November 26, 2014 and initiated April 24, 2014. Recalling firm: Medtronic Navigation, Inc.; Status: Terminated; Reported quantity: 26,674 total; Lot: Lot No.: D120065 D121495 D121515 D120175 D122611 D101555. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Medtronic Navigation, Inc.
Distributor
Nationwide Distribution
Lot
Lot No.: D120065 D121495 D121515 D120175 D122611 D101555
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was DRAPE 9732722 TUBE STERILE O-ARM 20PK recalled?

Yes. FDA published recall Z-1583-2014 on November 26, 2014. Reported hazard: Medtronic Navigation is recalling certain Medtronic single-use, disposable devices because there was not sufficient control to assure that returned, unused devices were appropriately examined for product and/or package damage to ensure sterility.