Maquet Getinge-BEQ-TOP 5208 ECC SET CARDIAC Material: 701067350

Published June 5, 2019 · Recall initiated: April 18, 2019

Summary

FDA records medical device Class II recall Z-1585-2019, published June 5, 2019 and initiated April 18, 2019. Recalling firm: Maquet Cardiovascular, LLC; Status: Terminated; Reported quantity: N/A; Lot: Batch Numbers: 3000044413 3000049064 3000050885 3000054157 3000058832 3000062461 3000065084. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Maquet Cardiovascular, LLC
Distributor
Nationwide
Lot
Batch Numbers: 3000044413 3000049064 3000050885 3000054157 3000058832 3000062461 3000065084
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Maquet Getinge-BEQ-TOP 5208 ECC SET CARDIAC Material: 701067350 recalled?

Yes. FDA published recall Z-1585-2019 on June 5, 2019. Reported hazard: Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured