Back to search
FDAFDA · Food and drugs·Recall number Z-1587-2019

Maquet Getinge-BEQ-T 37305 1/4+3/8 TABLE LINE Material: 701067577

Published June 5, 2019 · Recall initiated: April 18, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Maquet Cardiovascular, LLC
Distributor
Nationwide
Lot
Batch Numbers: 3000066690 3000072826 3000080852
Risk classification
Class II
Area (FDA)
device
Country
United States