FDAFDA · Food and drugs·Recall number Z-1595-2024

SW Bari Lift & Transfer, Model Number 38060000

Published April 24, 2024 · Recall initiated: January 25, 2024

Summary

FDA records medical device Class II recall Z-1595-2024, published April 24, 2024 and initiated January 25, 2024. Recalling firm: Raye's Inc.; Status: Ongoing; Reported quantity: 1,149 units; Lot: UDI/DI 00845699001040, All serial numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Raye's Inc.
Distributor
Worldwide distribution - US Nationwide and the countries of Australia, UK.
Lot
UDI/DI 00845699001040, All serial numbers
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was SW Bari Lift & Transfer, Model Number 38060000 recalled?

Yes. FDA published recall Z-1595-2024 on April 24, 2024. Reported hazard: Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.