SW Bari Lift & Transfer, Model Number 38060000
Published April 24, 2024 · Recall initiated: January 25, 2024
Summary
FDA records medical device Class II recall Z-1595-2024, published April 24, 2024 and initiated January 25, 2024. Recalling firm: Raye's Inc.; Status: Ongoing; Reported quantity: 1,149 units; Lot: UDI/DI 00845699001040, All serial numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Raye's Inc.
- Distributor
- Worldwide distribution - US Nationwide and the countries of Australia, UK.
- Lot
- UDI/DI 00845699001040, All serial numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMicroalbumin Calibrator Series (mALB CAL)-IVD to calibrate the Microalbumin Assay Catalog Number: MA2426April 24, 2024
- FDARENASYS EDGE 300ML CANISTER WITH Solidifier- Indicated for patients who would benefit from a suction pumpApril 24, 2024
- FDAMicroalbumin Calibrator Series (mALB CAL)-IVD used to calibrate the Microalbumin Assay Catalog Number: MA1567April 24, 2024
- FDARENASYS EDGE 800ML CANISTER WITH Solidifier- Indicated for patients who would benefit from a suction pumpApril 24, 2024
- FDAActim PROM dipstik, REF 30831ETUS; qualitative immunoassay rapid test for rupture of fetal membraneApril 24, 2024
- FDASiemens Artis Models: Interventional Fluoroscopic X-Ray Systems-Models: Artis zee floor- 10094135 Artis zeeApril 24, 2024
Questions about this recall
Was SW Bari Lift & Transfer, Model Number 38060000 recalled?
Yes. FDA published recall Z-1595-2024 on April 24, 2024. Reported hazard: Retrospective engineering analysis against current ISO standards for hoists has determined: 1. The device cannot physically perform at maximum documented weight specifications when a factor of safety of two times the maximum load is applied. 2. The device is not equipped with a safety device to ensure a person with disability would not fall in the event of a single-fault condition of the lifting device.
What should I do if I have this product?
Voluntary: Firm initiated · Letter