Back to search
FDAFDA · Food and drugs·Recall number Z-1596-2024

RENASYS EDGE 300ML CANISTER WITH Solidifier- Indicated for patients who would benefit from a suction pump (NPWT), as it may promote wound healing. Product Number: 66803139

Published April 24, 2024 · Recall initiated: March 11, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Black particulate matter from the carbon filter component of the canisters entering the NPWT pump and potentially causing damage to the pump

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Smith & Nephew Medical, Ltd.
Distributor
Worldwide distribution - US Nationwide and the countries of Canada, Chile.
Lot
GTIN: (01)05000223501044 Lot Numbers: 1016448, 1016478, 1016537, 1016773, 1017147, 1017417, 1017418, 1017912, 1016044
Risk classification
Class II
Area (FDA)
device
Country
United Kingdom

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card