Back to search
FDAFDA · Food and drugs·Recall number Z-1611-2019

Maquet Getinge-T 8030 4 in High Flow Pigtail Material: 709000366

Published June 5, 2019 · Recall initiated: April 18, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Maquet Cardiovascular, LLC
Distributor
Nationwide
Lot
Batch Numbers: 3000065493
Risk classification
Class II
Area (FDA)
device
Country
United States