Maquet Getinge-BEQ-T 4917 Mercy Medical Versa-Pack Material: 709000368
Published June 5, 2019 · Recall initiated: April 18, 2019
Summary
FDA records medical device Class II recall Z-1612-2019, published June 5, 2019 and initiated April 18, 2019. Recalling firm: Maquet Cardiovascular, LLC; Status: Terminated; Reported quantity: N/A; Lot: Batch Numbers: 3000043162. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Maquet Cardiovascular, LLC
- Distributor
- Nationwide
- Lot
- Batch Numbers: 3000043162
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThe Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile,July 22, 2026
- FDAVasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 EndoscopicOctober 1, 2025
- FDAAcrobat-i Positioner. Model Number C-XP-5000ZNovember 27, 2024
- FDAAcrobat SUV Vacuum Stabilizer System, ST. Model Number C-OM-9000ZNovember 27, 2024
- FDAMaquet Getinge-T 8030 4 in High Flow Pigtail Material: 709000366June 5, 2019
- FDAMaquet Getinge-BEQ-T 25501 3/8" Better Bladder Pack Material: 709000370June 5, 2019
Questions about this recall
Was Maquet Getinge-BEQ-T 4917 Mercy Medical Versa-Pack Material: 709000368 recalled?
Yes. FDA published recall Z-1612-2019 on June 5, 2019. Reported hazard: Custom Tubing Sets for Extracorporeal Circulation (ECC)-During sterilization, the closed stopcock/non-vented cap configuration may prevent the flow of humidity/ethylene oxide gas into the fluid-path between the closed port and the non-vented cap therefore the Sterility Assurance Level (SAL) of the stopcock ports cannot be assured
What should I do if I have this product?
Voluntary: Firm initiated · Letter