SYNCHRON System(s) Lactate (LACT) Reagent. Product Usage: Lactate (LACT) reagent, when used in conjunction with Synchron LX® System(s), UniCel® DxC 600/800 System(s) and Synchron® Systems Multi Calibrator, is intended for the quantitative determination of lactate concentration in human plasma and cerebrospinal fluid (CSF).
Published July 3, 2013 · Recall initiated: May 28, 2013
Summary
FDA records medical device Class II recall Z-1613-2013, published July 3, 2013 and initiated May 28, 2013. Recalling firm: Beckman Coulter Inc.; Status: Terminated; Reported quantity: 3,353 units; Lot: Part Number A95550, Lot Number M210077. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Beckman Coulter has confirmed customer reports of LACT Lot M210077 failing calibration with error condition OCR LOW. The calibration failures appear to be related to improper shipping or storage conditions (frozen reagent). Failed calibration of LACT may cause a delay in reporting results.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Beckman Coulter Inc.
- Distributor
- Worldwide Distribution - USA Nationwide and country of Canada.
- Lot
- Part Number A95550, Lot Number M210077
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADxC 500 AU Clinical Chemistry Analyzers: 1. AU w/out ISE, Part No. C63519 2. AU w/ISE, Part No. C63520October 14, 2026
- FDADxC 500i Clinical Analyzer Modules: 1. DxC 500 AU Module w/ISE, DxC 500i, Part No. C63522 2. DxC 500 AUOctober 14, 2026
- FDABeckman Coulter DxC 500i Clinical Analyzer Modules using software version 1.5 or lower: 1) DxC 500 AU ModuleAugust 26, 2026
- FDABeckman Coulter DxC 500 AU Clinical Analyzer Modules using software version 1.5 or lower: 1) AU w/out ISE,August 26, 2026
- FDATerumo Glidesheath (tm), 6 French, Nitinol wire, product code 15-1060. Used to facilitate placing a catheterJuly 3, 2013
- FDAVITROS Chemistry Products K+ (Potassium) Slides (Reagent), REF/Catalog Number 815 7596, Product Usage: For inJuly 3, 2013
Questions about this recall
Was SYNCHRON System(s) Lactate (LACT) Reagent. Product Usage: Lactate (LACT) reagent, when used in conjunction with Synchron LX® System(s), UniCel® DxC 600/800 System(s) and Synchron® Systems Multi Calibrator, is intended for the quantitative determination of lactate concentration in human plasma and cerebrospinal fluid (CSF). recalled?
Yes. FDA published recall Z-1613-2013 on July 3, 2013. Reported hazard: Beckman Coulter has confirmed customer reports of LACT Lot M210077 failing calibration with error condition OCR LOW. The calibration failures appear to be related to improper shipping or storage conditions (frozen reagent). Failed calibration of LACT may cause a delay in reporting results.
What should I do if I have this product?
Voluntary: Firm initiated · Letter