SYNCHRON System(s) Lactate (LACT) Reagent. Product Usage: Lactate (LACT) reagent, when used in conjunction with Synchron LX® System(s), UniCel® DxC 600/800 System(s) and Synchron® Systems Multi Calibrator, is intended for the quantitative determination of lactate concentration in human plasma and cerebrospinal fluid (CSF).

Published July 3, 2013 · Recall initiated: May 28, 2013

Summary

FDA records medical device Class II recall Z-1613-2013, published July 3, 2013 and initiated May 28, 2013. Recalling firm: Beckman Coulter Inc.; Status: Terminated; Reported quantity: 3,353 units; Lot: Part Number A95550, Lot Number M210077. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Beckman Coulter has confirmed customer reports of LACT Lot M210077 failing calibration with error condition OCR LOW. The calibration failures appear to be related to improper shipping or storage conditions (frozen reagent). Failed calibration of LACT may cause a delay in reporting results.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Beckman Coulter Inc.
Distributor
Worldwide Distribution - USA Nationwide and country of Canada.
Lot
Part Number A95550, Lot Number M210077
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was SYNCHRON System(s) Lactate (LACT) Reagent. Product Usage: Lactate (LACT) reagent, when used in conjunction with Synchron LX® System(s), UniCel® DxC 600/800 System(s) and Synchron® Systems Multi Calibrator, is intended for the quantitative determination of lactate concentration in human plasma and cerebrospinal fluid (CSF). recalled?

Yes. FDA published recall Z-1613-2013 on July 3, 2013. Reported hazard: Beckman Coulter has confirmed customer reports of LACT Lot M210077 failing calibration with error condition OCR LOW. The calibration failures appear to be related to improper shipping or storage conditions (frozen reagent). Failed calibration of LACT may cause a delay in reporting results.

What should I do if I have this product?

Voluntary: Firm initiated · Letter