VITROS Chemistry Products K+ (Potassium) Slides (Reagent), REF/Catalog Number 815 7596, Product Usage: For in vitro diagnostic use only. VITROS Chemistry Products K+ Slides quantitatively measure potassium (K+) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Published July 3, 2013 · Recall initiated: May 9, 2013
Summary
FDA records medical device Class III recall Z-1614-2013, published July 3, 2013 and initiated May 9, 2013. Recalling firm: Ortho-Clinical Diagnostics; Status: Terminated; Reported quantity: Domestic: 236,364; Foreign: 8,339; Lot: GEN 4102, Coating 0884, Lot XXXX (exp. 06/01/2014) and GEN 4102, Coating 0885, Lot XXXX (exp. 06/01/2014 and 07/01/2014). The term GEN and Coating refer to specific segments of the reagent lot number. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
While performing interval testing, Ortho Clinical Diagnostics (OCD) determined that a surfactant was unintentionally added to VITROS Chemistry Products K+ Slides during the manufacturing process.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Ortho-Clinical Diagnostics
- Distributor
- Worldwide Disribution - USA Nationwide including Puerto Rico, and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain and Venezuela.
- Lot
- GEN 4102, Coating 0884, Lot XXXX (exp. 06/01/2014) and GEN 4102, Coating 0885, Lot XXXX (exp. 06/01/2014 and 07/01/2014). The term GEN and Coating refer to specific segments of the reagent lot number.
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVITROS Performance Verifier IIApril 3, 2024
- FDAVITROS Performance Verifier IApril 3, 2024
- FDAVITROS XT Chemistry Products TBIL-ALKP Slides (ALKP is not affected) In Vitro Diagnostic use only - The TBILSeptember 2, 2020
- FDAVITROS Chemistry Product Performance Verifier I, Product Code 8231474, UDI # 10758750004577 - Product Usage:June 24, 2020
- FDASYNCHRON System(s) Lactate (LACT) Reagent. Product Usage: Lactate (LACT) reagent, when used in conjunctionJuly 3, 2013
- FDAVITROS Chemistry Products CRP Slides (Reagent), REF/Catalog Number 192 6740 and 809 7990, Product Usage: ForJuly 3, 2013
Questions about this recall
Was VITROS Chemistry Products K+ (Potassium) Slides (Reagent), REF/Catalog Number 815 7596, Product Usage: For in vitro diagnostic use only. VITROS Chemistry Products K+ Slides quantitatively measure potassium (K+) concentration in serum, plasma, and urine using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. recalled?
Yes. FDA published recall Z-1614-2013 on July 3, 2013. Reported hazard: While performing interval testing, Ortho Clinical Diagnostics (OCD) determined that a surfactant was unintentionally added to VITROS Chemistry Products K+ Slides during the manufacturing process.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit