FDAFDA · Food and drugs·Recall number Z-1615-2013
VITROS Chemistry Products CRP Slides (Reagent), REF/Catalog Number 192 6740 and 809 7990, Product Usage: For in vitro diagnostic use only. VITROS Chemistry Products CRP Slides quantitatively measure C-reactive protein (CRP) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.
Published July 3, 2013 · Recall initiated: May 9, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
While performing interval testing Ortho Clinical Diagnostics (OCD) determined that an excess amount of surfactant was unintentionally added to VITROS Chemistry Products CRP Slides during the manufacturing process.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Recall details
- Manufacturer
- Ortho-Clinical Diagnostics
- Distributor
- Worldwide Disribution - USA Nationwide including Puerto Rico, and the countries of Australia, Bermuda, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain and Venezuela.
- Lot
- GEN 3773, Coating 0766, Lot XXXX (exp. 11/01/2013); GEN 3784, Coating 0767, Lot XXXX (exp. 11/01/2013); and GEN 3785, Coating 0768, Lot XXXX (exp. 10/01/2013 and 11/01/2013). The term GEN and Coating refer to specific segments of the reagent lot number.
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States