FDAFDA · Food and drugs·Recall number Z-1627-2014
LATEX-FREE; 4-LEAD T-U-R IRRIGATION SET; Nonvented Set for Transurethral Resection Procedures; 85 inch (216 cm); For Flexible Irrigation Container Systems; List No. 15239-01; Single Use; Hospira Product Usage: For Transurethral Resection Procedures
Published May 28, 2014 · Recall initiated: March 25, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Tyvek cover may not be completely sealed to the rest of the package allowing openings at the edge of the sterile package for the following products: List No. 04693-01, Latex Free Extension Set; List No. 06543-01, Nonvented Y-Set for Transurethral Resection; List No. 06544-01 Cystoscopy / Irrigation Set; List No. 06599-01, Large Bore Y-Irrigation Set (Latex-Free); and, List No. 15239-01, 4-Lea
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Hospira Inc.
- Distributor
- Worldwide Distribution - US: Nationwide in the states of (AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, UT, WA, WI, WV, WY) including District of Columbia and Puerto Rico; *** the countries of Barbados, Canada, Costa Rica, Dominican Republic, Hong Kong, Kuwait, Libya, Saudi Arabia, United Arab Emirates
- Lot
- List No. 15239-04-01 (15239-01); Lot Numbers: 13259NS, 17185NS, 18181NS, 23102NS, 27108NS, 28275NS, 29093NS, 31041NS
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.