Hologic Selenia Full Field Digital Mammography System: SEL-00002, SEL-00013, SEL-00030, SEL-00039. The Selenia Full Field Digital Mammography system generates digital mammography images that can be used for screening and diagnosis of breast cancer.

Published June 11, 2014 · Recall initiated: May 13, 2014

Summary

FDA records medical device Class II recall Z-1700-2014, published June 11, 2014 and initiated May 13, 2014. Recalling firm: Hologic, Inc.; Status: Terminated; Reported quantity: 11 units; Lot: Serial Numbers: 2840314W8065, 2840314W8062W, 2840314W8066W, 2840314W8067 2840314W8073, 2840314W8064, 2840314W8056W, 28401082720 28407083579, 28401137611W, 28403083045W. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Hologic, Inc.
Distributor
Distributed in Puerto Rico, and the states of CA, NY, and TX, and the country of Republic of Korea.
Lot
Serial Numbers: 2840314W8065, 2840314W8062W, 2840314W8066W, 2840314W8067 2840314W8073, 2840314W8064, 2840314W8056W, 28401082720 28407083579, 28401137611W, 28403083045W
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Hologic Selenia Full Field Digital Mammography System: SEL-00002, SEL-00013, SEL-00030, SEL-00039. The Selenia Full Field Digital Mammography system generates digital mammography images that can be used for screening and diagnosis of breast cancer. recalled?

Yes. FDA published recall Z-1700-2014 on June 11, 2014. Reported hazard: Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury

What should I do if I have this product?

Voluntary: Firm initiated · Letter