Hologic M-IV Mammography System; Model M4--00001. Intended to produce radiographic images of the breast. For screening and diagnostic imaging of the breast.
Published June 11, 2014 · Recall initiated: May 13, 2014
Summary
FDA records medical device Class II recall Z-1701-2014, published June 11, 2014 and initiated May 13, 2014. Recalling firm: Hologic, Inc.; Status: Terminated; Reported quantity: 3 units; Lot: Serial Numbers: 1800314 7263, 180031472, 18002023051PR. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Hologic, Inc.
- Distributor
- Distributed in Puerto Rico, and the states of CA, NY, and TX, and the country of Republic of Korea.
- Lot
- Serial Numbers: 1800314 7263, 180031472, 18002023051PR
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThe Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN Specimen Collection kitFebruary 12, 2025
- FDAPanther Fusion SARS-CoV-2/Flu A/B/RSV Assay, REF: PRD-07400January 1, 2025
- FDAAptima CMV Quant Assay, Reference numbers: AW-22600-001, AW-25509-001, and AW-26381-001 Amplification Reagent,April 3, 2024
- FDAPANTHER FUSION System REF 9121010000 The Panther System is an integrated nucleic acid testing system whichJanuary 31, 2024
- FDAHologic Selenia Full Field Digital Mammography System: SEL-00002, SEL-00013, SEL-00030, SEL-00039. The SeleniaJune 11, 2014
- FDALifeCare 5000 Pump (Plum 1.6) infusion pump, indicated for volumetric infusion of critical drugs and fluids byJune 11, 2014
Questions about this recall
Was Hologic M-IV Mammography System; Model M4--00001. Intended to produce radiographic images of the breast. For screening and diagnostic imaging of the breast. recalled?
Yes. FDA published recall Z-1701-2014 on June 11, 2014. Reported hazard: Selenia Full Field Digital Mammography System defective relay switch, if there is a need to use the emergency off switch, the system may not turn off which could result in patient injury
What should I do if I have this product?
Voluntary: Firm initiated · Letter