FDAFDA · Food and drugs·Recall number Z-1725-2014

Leica Ceiling Mount MSF-1 for use with M841 and M501 operating microscopes. Article Number 10446781 utilizes a supply voltage of 230 V. Article Number 10446782 utilizes a supply voltage of 110 V. Intended to be used for improving the visibility of objects through magnification and illumination.

Published June 18, 2014 · Recall initiated: September 11, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6 inches).

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Leica Microsystems, Inc.
Distributor
Worldwide Distribution -- USA, including the states of AZ, IL, MA, MO, NC, OR, TX, VA, and WI; and, the countries of Australia, Austria, Chile, China, Colombia, Czech Republic, Greece, Israel, Italy, Japan, Malaysia, Netherlands, Norway, Oman, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, and United Kingdom.
Lot
Leica Ceiling Mount MSF-1; Article Numbers: 10446781, 10446782
Risk classification
Class II
Area (FDA)
device
Country
United States