Leica Ceiling Mount MS-1F & MC-1F for use with the M500 N operating microscope. Article Number 10446657 utilizes a supply voltage of 110 V. Article Number 10446658 utilizes a supply voltage of 220 V. Intended to be used for improving the visibility of objects through magnification and illumination.

Published June 18, 2014 · Recall initiated: September 11, 2013

Summary

FDA records medical device Class II recall Z-1726-2014, published June 18, 2014 and initiated September 11, 2013. Recalling firm: Leica Microsystems, Inc.; Status: Terminated; Reported quantity: 103 total; Lot: Leica Ceiling Mounts MS-1F, MC-1F; Article Numbers: 10446657, 10446658. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6 inches).

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Leica Microsystems, Inc.
Distributor
Worldwide Distribution -- USA, including the states of AZ, IL, MA, MO, NC, OR, TX, VA, and WI; and, the countries of Australia, Austria, Chile, China, Colombia, Czech Republic, Greece, Israel, Italy, Japan, Malaysia, Netherlands, Norway, Oman, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, and United Kingdom.
Lot
Leica Ceiling Mounts MS-1F, MC-1F; Article Numbers: 10446657, 10446658
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Leica Ceiling Mount MS-1F & MC-1F for use with the M500 N operating microscope. Article Number 10446657 utilizes a supply voltage of 110 V. Article Number 10446658 utilizes a supply voltage of 220 V. Intended to be used for improving the visibility of objects through magnification and illumination. recalled?

Yes. FDA published recall Z-1726-2014 on June 18, 2014. Reported hazard: The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6 inches).

What should I do if I have this product?

Voluntary: Firm initiated · Letter