Leica Ceiling Mount MS-1F & MC-1F for use with the M500 N operating microscope. Article Number 10446657 utilizes a supply voltage of 110 V. Article Number 10446658 utilizes a supply voltage of 220 V. Intended to be used for improving the visibility of objects through magnification and illumination.
Published June 18, 2014 · Recall initiated: September 11, 2013
Summary
FDA records medical device Class II recall Z-1726-2014, published June 18, 2014 and initiated September 11, 2013. Recalling firm: Leica Microsystems, Inc.; Status: Terminated; Reported quantity: 103 total; Lot: Leica Ceiling Mounts MS-1F, MC-1F; Article Numbers: 10446657, 10446658. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6 inches).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Leica Microsystems, Inc.
- Distributor
- Worldwide Distribution -- USA, including the states of AZ, IL, MA, MO, NC, OR, TX, VA, and WI; and, the countries of Australia, Austria, Chile, China, Colombia, Czech Republic, Greece, Israel, Italy, Japan, Malaysia, Netherlands, Norway, Oman, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, and United Kingdom.
- Lot
- Leica Ceiling Mounts MS-1F, MC-1F; Article Numbers: 10446657, 10446658
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABOND Ready-To-Use Primary Antibody CDX2 (EP25), REF PA0375September 22, 2021
- FDABOND Polymer Refine Detection Kit, Catalog No. DS9800. This detection system is for in vitro diagnostic use.December 18, 2019
- FDALeica M530 OHX Microscope System Supplied with 100 - 120 Vac.August 15, 2018
- FDANussloch GmbH ASP6025 Tissue ProcessorJuly 12, 2017
- FDALeica Ceiling Mount MSF-1 for use with M841 and M501 operating microscopes. Article Number 10446781 utilizes aJune 18, 2014
- FDAAlere Triage® D-Dimer Test PN 98100, Lot W53884B. The Alere Triage® D-Dimer Test is a fluorescence immunoassayJune 18, 2014
Questions about this recall
Was Leica Ceiling Mount MS-1F & MC-1F for use with the M500 N operating microscope. Article Number 10446657 utilizes a supply voltage of 110 V. Article Number 10446658 utilizes a supply voltage of 220 V. Intended to be used for improving the visibility of objects through magnification and illumination. recalled?
Yes. FDA published recall Z-1726-2014 on June 18, 2014. Reported hazard: The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6 inches).
What should I do if I have this product?
Voluntary: Firm initiated · Letter