FDAFDA · Food and drugs·Recall number Z-1726-2014
Leica Ceiling Mount MS-1F & MC-1F for use with the M500 N operating microscope. Article Number 10446657 utilizes a supply voltage of 110 V. Article Number 10446658 utilizes a supply voltage of 220 V. Intended to be used for improving the visibility of objects through magnification and illumination.
Published June 18, 2014 · Recall initiated: September 11, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The ceiling mount supplier encountered an event in which the drive screw in the motor unit, which lifts and lowers the lower lift arm, broke. As a result, the lift arm fell down approx. 60 cm (23.6 inches).
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Leica Microsystems, Inc.
- Distributor
- Worldwide Distribution -- USA, including the states of AZ, IL, MA, MO, NC, OR, TX, VA, and WI; and, the countries of Australia, Austria, Chile, China, Colombia, Czech Republic, Greece, Israel, Italy, Japan, Malaysia, Netherlands, Norway, Oman, Saudi Arabia, South Korea, Spain, Sweden, Switzerland, and United Kingdom.
- Lot
- Leica Ceiling Mounts MS-1F, MC-1F; Article Numbers: 10446657, 10446658
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?
Recalert watches your whole catalog for recalls like this one, automatically.