Vertical Spring Arm for Single or Dual Monitor Mount (Harmony LL 500 and Harmony LL 700 surgical lighting systems), Steris Corporation. Model numbers YG19051 and YG19262.
Published July 2, 2014 · Recall initiated: April 4, 2014
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for the welded seam in affected monitor spring arms to crack and/or break.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Steris Corporation
- Distributor
- Distribution to US (states: MA, NY, OH, PA & TX) and Canada.
- Lot
- Model #: YG19051, Serial #'s: 01108340587, 01117338881, 011106337516, 011106337525, 011106337527, 011106337534, 011106337535, 011106337538, 011106337539, 011106337541, 011106337626, 011107338878, 011107338882, 011107338883, 011108340587, 011108340600, 0111073388883 & 011108340600A. Model #: YG19262, Serial #: ART00001 - ART00010, CADS0001 - CADS0015, CBX0001 - CBX0010 & CXSG0008 - CXSG0022.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States