Total Knee Arthroplasty (TKA) 2, Total Knee Arthroplasty (TKA) 1, Partial Knee Arthroplasty (PKA) 3, Total Hip Arthroplasty (THA) 4.0, 4.1 on Mako 3.0, 3.1 Software: TKA 2.0.de.1+TKA 1.0.1.de.1+PKA3.0.2.de. Application Part Number: 700002190743-03

Published June 5, 2024 · Recall initiated: April 15, 2024

Summary

FDA records medical device Class II recall Z-1914-2024, published June 5, 2024 and initiated April 15, 2024. Recalling firm: Stryker Orthopaedics; Status: Ongoing; Reported quantity: 40 units; Lot: UDI-DI: (01)07613327629132(10). Software: TKA 2.0.de.1+TKA 1.0.1.de.1+PKA3.0.2.de. Application Part Number: 700002190743-03. Revision: AA, AB, AC, & AD. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Application software intended to be deployed on Mako 3.0, part number 209999, and Mako 3.1, part number 219999, observed an increase in the Software Error #3 (SE3) error code when a Mako System shutdown or a Mako System restart is not performed prior to switching between applications (i.e. TKA to THA), resulting in delay in treatment

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Stryker Orthopaedics
Distributor
Worldwide - US Nationwide and the countries of Australia, Austria, Belgium, Canada, Chile, China, Columbia, Denmark, Finland, France, Germany, Greece, India, Israel, Italy, Japan, Korea, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Puerto Rico, Qatar, Romania, Saudi Arabia, Singapore, Spain, South Africa, Sweden, Switzerland, Thailand, Turkey, UAE, and UK
Lot
UDI-DI: (01)07613327629132(10). Software: TKA 2.0.de.1+TKA 1.0.1.de.1+PKA3.0.2.de. Application Part Number: 700002190743-03. Revision: AA, AB, AC, & AD.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Total Knee Arthroplasty (TKA) 2, Total Knee Arthroplasty (TKA) 1, Partial Knee Arthroplasty (PKA) 3, Total Hip Arthroplasty (THA) 4.0, 4.1 on Mako 3.0, 3.1 Software: TKA 2.0.de.1+TKA 1.0.1.de.1+PKA3.0.2.de. Application Part Number: 700002190743-03 recalled?

Yes. FDA published recall Z-1914-2024 on June 5, 2024. Reported hazard: Application software intended to be deployed on Mako 3.0, part number 209999, and Mako 3.1, part number 219999, observed an increase in the Software Error #3 (SE3) error code when a Mako System shutdown or a Mako System restart is not performed prior to switching between applications (i.e. TKA to THA), resulting in delay in treatment

What should I do if I have this product?

Voluntary: Firm initiated · Letter