epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;
Published May 6, 2026 · Recall initiated: April 2, 2026
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics Inc
- Distributor
- Worldwide distribution - US Nationwide and the countries of Italy, Spain, Austria, Portugal, Ireland, France, Romania, Canada, India, Philippines, Mexico.
- Lot
- Siemens Material Number (SMN): 10736515; UDI-DI: 00809708121860; Lot numbers: 01-26024-70, 01-26026-70, 01-26028-70, 01-26030-70 (OUS only), 02-26027-80, 02-26028-80 (OUS only);
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States