RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.

Published August 6, 2025 · Recall initiated: June 26, 2025

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Healthcare Diagnostics Inc
Distributor
worldwide distribution - US Nationwide and the countries of Canada, Chile, Japan, Paraguay.
Lot
Material Number: 10329097. GTIN Number: 00630414482163. Lot Number: WW/08925. Expiration Date: 12/03/2025.
Risk classification
Class II
Area (FDA)
device
Country
United States