RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097.
Published August 6, 2025 · Recall initiated: June 26, 2025
Summary
FDA records medical device Class II recall Z-2230-2025, published August 6, 2025 and initiated June 26, 2025. Recalling firm: Siemens Healthcare Diagnostics Inc; Status: Ongoing; Reported quantity: 1,477 units; Lot: Material Number: 10329097. GTIN Number: 00630414482163. Lot Number: WW/08925. Expiration Date: 12/03/2025. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics Inc
- Distributor
- worldwide distribution - US Nationwide and the countries of Canada, Chile, Japan, Paraguay.
- Lot
- Material Number: 10329097. GTIN Number: 00630414482163. Lot Number: WW/08925. Expiration Date: 12/03/2025.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAepoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;May 6, 2026
- FDAIngenia Elition X Model Numbers (REF): (1) 781358, (2) 782107, (3) 782119, (4) 782136, (5) 782151 (ChinaAugust 6, 2025
- FDA3M Ranger Irrigation Fluid Warming Set, REF 24750August 6, 2025
- FDAepoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System, SMN 10736515November 13, 2024
- FDASoftware version 5.3 in: RAPIDPoint 500e Blood Gas System (USA) Siemens Material Number: 11416751 RAPIDPointJuly 3, 2024
- FDARAPIDPoint® 500 Blood Gas Analyzer with v2.4 software for professional use in a point-of-care or laboratoryMay 17, 2017
Questions about this recall
Was RAPIDPoint 500 Systems Wash/Waste Cartridge (4 Pack). Material Number: 10329097. recalled?
Yes. FDA published recall Z-2230-2025 on August 6, 2025. Reported hazard: Siemens Healthcare Diagnostics is recalling the RAPIDPoint 500 Systems Wash/Waste Cartridge Lot WW/08925 due to a manufacturing error causing positive and negative bias in Ionized Calcium, Potassium, pH, Chloride, pCO2, Glucose and Lactate analytes and their derivatives. Four complaints were filed related to this event that led to Siemens Healthcare Diagnostics identifying an error in the manufacturing process and initiated corrective actions. Use of the affected device may result in errors during QC which could lead to a delay in patient treatment. If bias generated from this issue is not caught during QC erroneous results may cause temporary or medically reversible adverse health consequences to patients.
What should I do if I have this product?
Voluntary: Firm initiated · Letter