Reliance 1227 Cart and Utensil Washer/Disinfector Product Usage: Usage: The Reliance 1227 Cart and Utensil Washer/Disinfector is intended for use in the cleaning and low-level disinfection of bedpans and urinals, basins, carts, beds, theater shoes and other miscellaneous reusable items used in the care of patients.

Published July 16, 2014 · Recall initiated: June 30, 2014

Summary

FDA records medical device Class III recall Z-2036-2014, published July 16, 2014 and initiated June 30, 2014. Recalling firm: Steris Corporation; Status: Terminated; Reported quantity: 37; Lot: Serial Number Range: 3632511020 - 3613214010. Class III means the product is not likely to cause adverse health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

STERIS has identified that deformation of the washers door handle pin may occur when the door handle is rotated with sufficient force. Should deformation and/or breakage of the door handle pin occur there is a potential for the washer door to become inoperable.

What to do

Voluntary: Firm initiated · Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Steris Corporation
Distributor
US Nationwide Distribution - in the states of AL, AR, CA, DE, FL, GA, IL, IA, LA, MA, MI, NJ, NY, OH, OK, OR, SC, TX, VA. and in the country of Canada.
Lot
Serial Number Range: 3632511020 - 3613214010
Risk classification
Class III
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Reliance 1227 Cart and Utensil Washer/Disinfector Product Usage: Usage: The Reliance 1227 Cart and Utensil Washer/Disinfector is intended for use in the cleaning and low-level disinfection of bedpans and urinals, basins, carts, beds, theater shoes and other miscellaneous reusable items used in the care of patients. recalled?

Yes. FDA published recall Z-2036-2014 on July 16, 2014. Reported hazard: STERIS has identified that deformation of the washers door handle pin may occur when the door handle is rotated with sufficient force. Should deformation and/or breakage of the door handle pin occur there is a potential for the washer door to become inoperable.

What should I do if I have this product?

Voluntary: Firm initiated · Visit