OEC 6800 Miniview, Image-intensified fluoroscopic x-ray system used during diagnostic, surgical and interventional procedures. Designed to provide the physician with general fluoroscopic visualization of the patient including but not limited to surgical orthopedic and extremity imagine. The device is not intended for whole-body pediatric imaging.

Published August 1, 2012 · Recall initiated: November 20, 2006

Summary

FDA records medical device Class I recall Z-2040-2012, published August 1, 2012 and initiated November 20, 2006. Recalling firm: GE OEC Medical Systems, Inc; Status: Terminated; Reported quantity: 1,933 units; Lot: 86-1834 , 86-1175 , 86-0238 , 86-0585 , 86-1230 , 86-1992 , 86-0184 , 86-1033 , 86-1480 , 86-1369 , 86-0579 , 86-1678 , 86-1059 , 86-0639 , 86-0389 & 86-0390 , 86-0775 , 86-0850 & 86-1703 , 86-0151 & 86-1311 , 86-1883 , 86-0048 & 86-2015 , 86-0915, 86-1626 & 86-1631 , 86-2118 , 86-0949 , 86-0937 , 86-0071 & 86-1015 , 86-0352 , 86-0527 , 86-0140 , 86-1697 , 86-0697 , 86-1162 , 86-1424 , 86-0795 , 86-0260 , 86-0077 , 86-2002 , 86-0811 , 86-2148 , 86-1582 , 86-1633 , 86-0454 , 86-0185, 86-1119 & 86-1885 , 86-0668 , 86-1026 , 86-0529 , 86-1754 , 86-0901 , 86-0228 , 86-1555 , 86-1212 , 86-1095 , 86-0997 , 86-1475 , 86-2137 , 86-0829 , 86-1737 , 86-1259 , 86-0013 , 86-1139 , 86-0058 , 86-0327 , 86-1346 , 86-0419 & 86-2047 , 86-1892 , 86-0359 & 86-1070 , 86-0331 , 86-1255 , 86-0785 , 86-1043 , 86-1854 , 86-0148 & 86-1036 , 86-0427 & 86-2068 , 86-1988 , 86-0092 , 86-1283 , 86-1955 , 86-0711 , 86-2035 , 86-2017 , 86-0770 & 86-2044 , 86-0716 , 86-0692 , 86-1610 , 86-0154 & 86-1822 , 86-1727 , 86-1209 & 86-1210 , 86-0046 , 86-1303 , 86-0468 , 86-1274 , 86-0113 , 86-0035 , 86-1481 , 86-1808 , 86-1332 , 86-0618 , 86-1860 , 86-0425 , 86-0912 , 86-0925 , 86-0306 & 86-0664 , 86-0987 , 86-2170 , 86-0308 , 86-1302 , 86-0399 , 86-1205 , 86-1615 , 86-2128 , 86-0162 , 86-0166 , 86-0105 , 86-1687 & 86-1844 , 86-0865 , 86-1045 , 86-1360 , 86-0903 , 86-1456 , 86-1213 , 86-1326 , 86-0745 , 86-0517 , 86-2140 , 86-0625 & 86-1399 , 86-0923 , 86-1319 & 86-2133 , 86-1499 , 86-1823 , 86-0401 , 86-0157 , 86-0685 , 86-0570 , 86-1446 , 86-1606 , 86-2030 , 86-0086 , 86-0075 & 86-0076 , 86-0435 , 86-1780 , 86-0247-RC , 86-1478 , 86-1894 , 86-0896 , 86-1249 , 86-0601, 86-0914 & 86-1248 , 86-1250 , 86-1836 , 86-0805, 86-0806 & 86-0972 , 86-1619 , 86-1576 , 86-0544 , 86-1886 , 86-1782 , 86-0475 , 86-0321 , 86-0511 , 86-1949 , 86-1150 , 86-1065 , 86-1614 , 86-1907 , 86-1625 , 86-0757 & 86-0786 , 86-2092 , 86-0501 , 86-2117 , 86-1879 , 86-1714 , 86-0416 , 86-1506 , 86-1732 , 86-0977 , 86-0964 , 86-1531 , 86-0994 & 86-1909 , 86-0091 , 86-0115 , 86-1154 , 86-2171 , 86-0739 , 86-1621 & 86-1721 , 86-1647 , 86-0589 , 86-1490 , 86-0485 , 86-0995 , 86-0657 , 86-0700 , 86-1620 , 86-2032 , 86-0487 & 86-0494 , 86-1343 , 86-0102 , 86-2005 , 86-0504 , 86-1178 , 86-0329 , 86-0777 & 86-1273 , 86-1187 , 86-0519 , 86-1536 , 86-1448 , 86-0572 , 86-1223 , 86-2199 , 86-1880 , 86-0827 , 86-0068, 86-0155 & 86-0239 , 86-0492 , 86-0558 & 86-1966 , 86-0347 & 86-0411 , 86-1324 , 86-0191 , 86-1166 , 86-0701 , 86-2008 , 86-1219 , 86-0737 , 86-0974 , 86-2095 , 86-1460 , 86-0381 , 86-1234 , 86-0890 , 86-0099 , 86-1420 , 86-0750 , 86-0259 , 86-1226 , 86-0204 , 86-0482 , 86-0378 , 86-2194 , 86-1293 , 86-0966 , 86-1111 , 86-1775 & 86-1965 , 86-0069 , 86-0137, 86-0428 & 86-0432 , 86-1813 , 86-0978 , 86-0609 , 86-0047 , 86-0070 , 86-1646 & 86-1838 , 86-0906 , 86-1432 , 86-1391 , 86-1431 , 86-1266 & 86-2238 , 86-2239 , 86-1375 , 86-0560 , 86-1686 , 86-1218 , 86-1308 , 86-0277 , 86-0755 & 86-1959 , 86-1179 , 86-0245 & 86-0253 , 86-2056 , 86-1151 , 86-2043 , 86-1192 & 86-1632 , 86-0840 , 86-1728 , 86-0025 , 86-1022 , 86-0871 , 86-0163 , 86-0271 , 86-1962 , 86-0153 & 86-1267 , 86-1523 & 86-1560 , 86-1611 , 86-0732 , 86-1540 , 86-1094 , 86-0573 & 86-1211 , 86-1494 , 86-0198 , 86-0388 , 86-0200 & 86-1381 , 86-2087 , 86-0981 & 86-1713 , 86-0120 , 86-1914 , 86-1972 , 86-1918 , 86-0106 , 86-1858 & 86-1889 , 86-1869 , 86-1583 , 86-1085. Class I means a reasonable probability of serious adverse health consequences or death.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. The OEC 6800 Miniview was recalled in November 2006 due to several intermittent potential safety issues including navigation tracking inaccuracy, system Lock-up, incorrect image display, incorrect dose area calculated air kerma in "film mo

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
GE OEC Medical Systems, Inc
Distributor
Worldwide Distribution
Lot
86-1834 , 86-1175 , 86-0238 , 86-0585 , 86-1230 , 86-1992 , 86-0184 , 86-1033 , 86-1480 , 86-1369 , 86-0579 , 86-1678 , 86-1059 , 86-0639 , 86-0389 & 86-0390 , 86-0775 , 86-0850 & 86-1703 , 86-0151 & 86-1311 , 86-1883 , 86-0048 & 86-2015 , 86-0915, 86-1626 & 86-1631 , 86-2118 , 86-0949 , 86-0937 , 86-0071 & 86-1015 , 86-0352 , 86-0527 , 86-0140 , 86-1697 , 86-0697 , 86-1162 , 86-1424 , 86-0795 , 86-0260 , 86-0077 , 86-2002 , 86-0811 , 86-2148 , 86-1582 , 86-1633 , 86-0454 , 86-0185, 86-1119 & 86-1885 , 86-0668 , 86-1026 , 86-0529 , 86-1754 , 86-0901 , 86-0228 , 86-1555 , 86-1212 , 86-1095 , 86-0997 , 86-1475 , 86-2137 , 86-0829 , 86-1737 , 86-1259 , 86-0013 , 86-1139 , 86-0058 , 86-0327 , 86-1346 , 86-0419 & 86-2047 , 86-1892 , 86-0359 & 86-1070 , 86-0331 , 86-1255 , 86-0785 , 86-1043 , 86-1854 , 86-0148 & 86-1036 , 86-0427 & 86-2068 , 86-1988 , 86-0092 , 86-1283 , 86-1955 , 86-0711 , 86-2035 , 86-2017 , 86-0770 & 86-2044 , 86-0716 , 86-0692 , 86-1610 , 86-0154 & 86-1822 , 86-1727 , 86-1209 & 86-1210 , 86-0046 , 86-1303 , 86-0468 , 86-1274 , 86-0113 , 86-0035 , 86-1481 , 86-1808 , 86-1332 , 86-0618 , 86-1860 , 86-0425 , 86-0912 , 86-0925 , 86-0306 & 86-0664 , 86-0987 , 86-2170 , 86-0308 , 86-1302 , 86-0399 , 86-1205 , 86-1615 , 86-2128 , 86-0162 , 86-0166 , 86-0105 , 86-1687 & 86-1844 , 86-0865 , 86-1045 , 86-1360 , 86-0903 , 86-1456 , 86-1213 , 86-1326 , 86-0745 , 86-0517 , 86-2140 , 86-0625 & 86-1399 , 86-0923 , 86-1319 & 86-2133 , 86-1499 , 86-1823 , 86-0401 , 86-0157 , 86-0685 , 86-0570 , 86-1446 , 86-1606 , 86-2030 , 86-0086 , 86-0075 & 86-0076 , 86-0435 , 86-1780 , 86-0247-RC , 86-1478 , 86-1894 , 86-0896 , 86-1249 , 86-0601, 86-0914 & 86-1248 , 86-1250 , 86-1836 , 86-0805, 86-0806 & 86-0972 , 86-1619 , 86-1576 , 86-0544 , 86-1886 , 86-1782 , 86-0475 , 86-0321 , 86-0511 , 86-1949 , 86-1150 , 86-1065 , 86-1614 , 86-1907 , 86-1625 , 86-0757 & 86-0786 , 86-2092 , 86-0501 , 86-2117 , 86-1879 , 86-1714 , 86-0416 , 86-1506 , 86-1732 , 86-0977 , 86-0964 , 86-1531 , 86-0994 & 86-1909 , 86-0091 , 86-0115 , 86-1154 , 86-2171 , 86-0739 , 86-1621 & 86-1721 , 86-1647 , 86-0589 , 86-1490 , 86-0485 , 86-0995 , 86-0657 , 86-0700 , 86-1620 , 86-2032 , 86-0487 & 86-0494 , 86-1343 , 86-0102 , 86-2005 , 86-0504 , 86-1178 , 86-0329 , 86-0777 & 86-1273 , 86-1187 , 86-0519 , 86-1536 , 86-1448 , 86-0572 , 86-1223 , 86-2199 , 86-1880 , 86-0827 , 86-0068, 86-0155 & 86-0239 , 86-0492 , 86-0558 & 86-1966 , 86-0347 & 86-0411 , 86-1324 , 86-0191 , 86-1166 , 86-0701 , 86-2008 , 86-1219 , 86-0737 , 86-0974 , 86-2095 , 86-1460 , 86-0381 , 86-1234 , 86-0890 , 86-0099 , 86-1420 , 86-0750 , 86-0259 , 86-1226 , 86-0204 , 86-0482 , 86-0378 , 86-2194 , 86-1293 , 86-0966 , 86-1111 , 86-1775 & 86-1965 , 86-0069 , 86-0137, 86-0428 & 86-0432 , 86-1813 , 86-0978 , 86-0609 , 86-0047 , 86-0070 , 86-1646 & 86-1838 , 86-0906 , 86-1432 , 86-1391 , 86-1431 , 86-1266 & 86-2238 , 86-2239 , 86-1375 , 86-0560 , 86-1686 , 86-1218 , 86-1308 , 86-0277 , 86-0755 & 86-1959 , 86-1179 , 86-0245 & 86-0253 , 86-2056 , 86-1151 , 86-2043 , 86-1192 & 86-1632 , 86-0840 , 86-1728 , 86-0025 , 86-1022 , 86-0871 , 86-0163 , 86-0271 , 86-1962 , 86-0153 & 86-1267 , 86-1523 & 86-1560 , 86-1611 , 86-0732 , 86-1540 , 86-1094 , 86-0573 & 86-1211 , 86-1494 , 86-0198 , 86-0388 , 86-0200 & 86-1381 , 86-2087 , 86-0981 & 86-1713 , 86-0120 , 86-1914 , 86-1972 , 86-1918 , 86-0106 , 86-1858 & 86-1889 , 86-1869 , 86-1583 , 86-1085
Risk classification
Class I
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was OEC 6800 Miniview, Image-intensified fluoroscopic x-ray system used during diagnostic, surgical and interventional procedures. Designed to provide the physician with general fluoroscopic visualization of the patient including but not limited to surgical orthopedic and extremity imagine. The device is not intended for whole-body pediatric imaging. recalled?

Yes. FDA published recall Z-2040-2012 on August 1, 2012. Reported hazard: Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. The OEC 6800 Miniview was recalled in November 2006 due to several intermittent potential safety issues including navigation tracking inaccuracy, system Lock-up, incorrect image display, incorrect dose area calculated air kerma in "film mo

What should I do if I have this product?

Voluntary: Firm initiated · Letter