OEC 6800 Miniview, Image-intensified fluoroscopic x-ray system used during diagnostic, surgical and interventional procedures. Designed to provide the physician with general fluoroscopic visualization of the patient including but not limited to surgical orthopedic and extremity imagine. The device is not intended for whole-body pediatric imaging.
Published August 1, 2012 · Recall initiated: November 20, 2006
Summary
FDA records medical device Class I recall Z-2040-2012, published August 1, 2012 and initiated November 20, 2006. Recalling firm: GE OEC Medical Systems, Inc; Status: Terminated; Reported quantity: 1,933 units; Lot: 86-1834 , 86-1175 , 86-0238 , 86-0585 , 86-1230 , 86-1992 , 86-0184 , 86-1033 , 86-1480 , 86-1369 , 86-0579 , 86-1678 , 86-1059 , 86-0639 , 86-0389 & 86-0390 , 86-0775 , 86-0850 & 86-1703 , 86-0151 & 86-1311 , 86-1883 , 86-0048 & 86-2015 , 86-0915, 86-1626 & 86-1631 , 86-2118 , 86-0949 , 86-0937 , 86-0071 & 86-1015 , 86-0352 , 86-0527 , 86-0140 , 86-1697 , 86-0697 , 86-1162 , 86-1424 , 86-0795 , 86-0260 , 86-0077 , 86-2002 , 86-0811 , 86-2148 , 86-1582 , 86-1633 , 86-0454 , 86-0185, 86-1119 & 86-1885 , 86-0668 , 86-1026 , 86-0529 , 86-1754 , 86-0901 , 86-0228 , 86-1555 , 86-1212 , 86-1095 , 86-0997 , 86-1475 , 86-2137 , 86-0829 , 86-1737 , 86-1259 , 86-0013 , 86-1139 , 86-0058 , 86-0327 , 86-1346 , 86-0419 & 86-2047 , 86-1892 , 86-0359 & 86-1070 , 86-0331 , 86-1255 , 86-0785 , 86-1043 , 86-1854 , 86-0148 & 86-1036 , 86-0427 & 86-2068 , 86-1988 , 86-0092 , 86-1283 , 86-1955 , 86-0711 , 86-2035 , 86-2017 , 86-0770 & 86-2044 , 86-0716 , 86-0692 , 86-1610 , 86-0154 & 86-1822 , 86-1727 , 86-1209 & 86-1210 , 86-0046 , 86-1303 , 86-0468 , 86-1274 , 86-0113 , 86-0035 , 86-1481 , 86-1808 , 86-1332 , 86-0618 , 86-1860 , 86-0425 , 86-0912 , 86-0925 , 86-0306 & 86-0664 , 86-0987 , 86-2170 , 86-0308 , 86-1302 , 86-0399 , 86-1205 , 86-1615 , 86-2128 , 86-0162 , 86-0166 , 86-0105 , 86-1687 & 86-1844 , 86-0865 , 86-1045 , 86-1360 , 86-0903 , 86-1456 , 86-1213 , 86-1326 , 86-0745 , 86-0517 , 86-2140 , 86-0625 & 86-1399 , 86-0923 , 86-1319 & 86-2133 , 86-1499 , 86-1823 , 86-0401 , 86-0157 , 86-0685 , 86-0570 , 86-1446 , 86-1606 , 86-2030 , 86-0086 , 86-0075 & 86-0076 , 86-0435 , 86-1780 , 86-0247-RC , 86-1478 , 86-1894 , 86-0896 , 86-1249 , 86-0601, 86-0914 & 86-1248 , 86-1250 , 86-1836 , 86-0805, 86-0806 & 86-0972 , 86-1619 , 86-1576 , 86-0544 , 86-1886 , 86-1782 , 86-0475 , 86-0321 , 86-0511 , 86-1949 , 86-1150 , 86-1065 , 86-1614 , 86-1907 , 86-1625 , 86-0757 & 86-0786 , 86-2092 , 86-0501 , 86-2117 , 86-1879 , 86-1714 , 86-0416 , 86-1506 , 86-1732 , 86-0977 , 86-0964 , 86-1531 , 86-0994 & 86-1909 , 86-0091 , 86-0115 , 86-1154 , 86-2171 , 86-0739 , 86-1621 & 86-1721 , 86-1647 , 86-0589 , 86-1490 , 86-0485 , 86-0995 , 86-0657 , 86-0700 , 86-1620 , 86-2032 , 86-0487 & 86-0494 , 86-1343 , 86-0102 , 86-2005 , 86-0504 , 86-1178 , 86-0329 , 86-0777 & 86-1273 , 86-1187 , 86-0519 , 86-1536 , 86-1448 , 86-0572 , 86-1223 , 86-2199 , 86-1880 , 86-0827 , 86-0068, 86-0155 & 86-0239 , 86-0492 , 86-0558 & 86-1966 , 86-0347 & 86-0411 , 86-1324 , 86-0191 , 86-1166 , 86-0701 , 86-2008 , 86-1219 , 86-0737 , 86-0974 , 86-2095 , 86-1460 , 86-0381 , 86-1234 , 86-0890 , 86-0099 , 86-1420 , 86-0750 , 86-0259 , 86-1226 , 86-0204 , 86-0482 , 86-0378 , 86-2194 , 86-1293 , 86-0966 , 86-1111 , 86-1775 & 86-1965 , 86-0069 , 86-0137, 86-0428 & 86-0432 , 86-1813 , 86-0978 , 86-0609 , 86-0047 , 86-0070 , 86-1646 & 86-1838 , 86-0906 , 86-1432 , 86-1391 , 86-1431 , 86-1266 & 86-2238 , 86-2239 , 86-1375 , 86-0560 , 86-1686 , 86-1218 , 86-1308 , 86-0277 , 86-0755 & 86-1959 , 86-1179 , 86-0245 & 86-0253 , 86-2056 , 86-1151 , 86-2043 , 86-1192 & 86-1632 , 86-0840 , 86-1728 , 86-0025 , 86-1022 , 86-0871 , 86-0163 , 86-0271 , 86-1962 , 86-0153 & 86-1267 , 86-1523 & 86-1560 , 86-1611 , 86-0732 , 86-1540 , 86-1094 , 86-0573 & 86-1211 , 86-1494 , 86-0198 , 86-0388 , 86-0200 & 86-1381 , 86-2087 , 86-0981 & 86-1713 , 86-0120 , 86-1914 , 86-1972 , 86-1918 , 86-0106 , 86-1858 & 86-1889 , 86-1869 , 86-1583 , 86-1085. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. The OEC 6800 Miniview was recalled in November 2006 due to several intermittent potential safety issues including navigation tracking inaccuracy, system Lock-up, incorrect image display, incorrect dose area calculated air kerma in "film mo
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE OEC Medical Systems, Inc
- Distributor
- Worldwide Distribution
- Lot
- 86-1834 , 86-1175 , 86-0238 , 86-0585 , 86-1230 , 86-1992 , 86-0184 , 86-1033 , 86-1480 , 86-1369 , 86-0579 , 86-1678 , 86-1059 , 86-0639 , 86-0389 & 86-0390 , 86-0775 , 86-0850 & 86-1703 , 86-0151 & 86-1311 , 86-1883 , 86-0048 & 86-2015 , 86-0915, 86-1626 & 86-1631 , 86-2118 , 86-0949 , 86-0937 , 86-0071 & 86-1015 , 86-0352 , 86-0527 , 86-0140 , 86-1697 , 86-0697 , 86-1162 , 86-1424 , 86-0795 , 86-0260 , 86-0077 , 86-2002 , 86-0811 , 86-2148 , 86-1582 , 86-1633 , 86-0454 , 86-0185, 86-1119 & 86-1885 , 86-0668 , 86-1026 , 86-0529 , 86-1754 , 86-0901 , 86-0228 , 86-1555 , 86-1212 , 86-1095 , 86-0997 , 86-1475 , 86-2137 , 86-0829 , 86-1737 , 86-1259 , 86-0013 , 86-1139 , 86-0058 , 86-0327 , 86-1346 , 86-0419 & 86-2047 , 86-1892 , 86-0359 & 86-1070 , 86-0331 , 86-1255 , 86-0785 , 86-1043 , 86-1854 , 86-0148 & 86-1036 , 86-0427 & 86-2068 , 86-1988 , 86-0092 , 86-1283 , 86-1955 , 86-0711 , 86-2035 , 86-2017 , 86-0770 & 86-2044 , 86-0716 , 86-0692 , 86-1610 , 86-0154 & 86-1822 , 86-1727 , 86-1209 & 86-1210 , 86-0046 , 86-1303 , 86-0468 , 86-1274 , 86-0113 , 86-0035 , 86-1481 , 86-1808 , 86-1332 , 86-0618 , 86-1860 , 86-0425 , 86-0912 , 86-0925 , 86-0306 & 86-0664 , 86-0987 , 86-2170 , 86-0308 , 86-1302 , 86-0399 , 86-1205 , 86-1615 , 86-2128 , 86-0162 , 86-0166 , 86-0105 , 86-1687 & 86-1844 , 86-0865 , 86-1045 , 86-1360 , 86-0903 , 86-1456 , 86-1213 , 86-1326 , 86-0745 , 86-0517 , 86-2140 , 86-0625 & 86-1399 , 86-0923 , 86-1319 & 86-2133 , 86-1499 , 86-1823 , 86-0401 , 86-0157 , 86-0685 , 86-0570 , 86-1446 , 86-1606 , 86-2030 , 86-0086 , 86-0075 & 86-0076 , 86-0435 , 86-1780 , 86-0247-RC , 86-1478 , 86-1894 , 86-0896 , 86-1249 , 86-0601, 86-0914 & 86-1248 , 86-1250 , 86-1836 , 86-0805, 86-0806 & 86-0972 , 86-1619 , 86-1576 , 86-0544 , 86-1886 , 86-1782 , 86-0475 , 86-0321 , 86-0511 , 86-1949 , 86-1150 , 86-1065 , 86-1614 , 86-1907 , 86-1625 , 86-0757 & 86-0786 , 86-2092 , 86-0501 , 86-2117 , 86-1879 , 86-1714 , 86-0416 , 86-1506 , 86-1732 , 86-0977 , 86-0964 , 86-1531 , 86-0994 & 86-1909 , 86-0091 , 86-0115 , 86-1154 , 86-2171 , 86-0739 , 86-1621 & 86-1721 , 86-1647 , 86-0589 , 86-1490 , 86-0485 , 86-0995 , 86-0657 , 86-0700 , 86-1620 , 86-2032 , 86-0487 & 86-0494 , 86-1343 , 86-0102 , 86-2005 , 86-0504 , 86-1178 , 86-0329 , 86-0777 & 86-1273 , 86-1187 , 86-0519 , 86-1536 , 86-1448 , 86-0572 , 86-1223 , 86-2199 , 86-1880 , 86-0827 , 86-0068, 86-0155 & 86-0239 , 86-0492 , 86-0558 & 86-1966 , 86-0347 & 86-0411 , 86-1324 , 86-0191 , 86-1166 , 86-0701 , 86-2008 , 86-1219 , 86-0737 , 86-0974 , 86-2095 , 86-1460 , 86-0381 , 86-1234 , 86-0890 , 86-0099 , 86-1420 , 86-0750 , 86-0259 , 86-1226 , 86-0204 , 86-0482 , 86-0378 , 86-2194 , 86-1293 , 86-0966 , 86-1111 , 86-1775 & 86-1965 , 86-0069 , 86-0137, 86-0428 & 86-0432 , 86-1813 , 86-0978 , 86-0609 , 86-0047 , 86-0070 , 86-1646 & 86-1838 , 86-0906 , 86-1432 , 86-1391 , 86-1431 , 86-1266 & 86-2238 , 86-2239 , 86-1375 , 86-0560 , 86-1686 , 86-1218 , 86-1308 , 86-0277 , 86-0755 & 86-1959 , 86-1179 , 86-0245 & 86-0253 , 86-2056 , 86-1151 , 86-2043 , 86-1192 & 86-1632 , 86-0840 , 86-1728 , 86-0025 , 86-1022 , 86-0871 , 86-0163 , 86-0271 , 86-1962 , 86-0153 & 86-1267 , 86-1523 & 86-1560 , 86-1611 , 86-0732 , 86-1540 , 86-1094 , 86-0573 & 86-1211 , 86-1494 , 86-0198 , 86-0388 , 86-0200 & 86-1381 , 86-2087 , 86-0981 & 86-1713 , 86-0120 , 86-1914 , 86-1972 , 86-1918 , 86-0106 , 86-1858 & 86-1889 , 86-1869 , 86-1583 , 86-1085
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAOEC 3D REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 3D is a mobile fluoroscopic C-ArmDecember 25, 2024
- FDAOEC 9800 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The Series OEC 9800 is designed to provideDecember 25, 2024
- FDAOEC 9900 REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC 9900 Elite mobile fluoroscopy systemDecember 25, 2024
- FDAOEC Elite REF Image-Intensified Fluoroscopic X-Ray System, Mobile. The OEC Elite mobile C-arm is intended toDecember 25, 2024
- FDAOEC 9800; OEC FluoroTrak 9800 Plus; OEC 9800 Plus; OEC 9800 MD Motorized C-arm System, 892.1650August 1, 2012
- FDALeica M525 F50 Surgical Microscope; Article No. 10338422; Leica Microsystems (Schweiz) Ag, 201 Max SchmidheinyAugust 1, 2012
Questions about this recall
Was OEC 6800 Miniview, Image-intensified fluoroscopic x-ray system used during diagnostic, surgical and interventional procedures. Designed to provide the physician with general fluoroscopic visualization of the patient including but not limited to surgical orthopedic and extremity imagine. The device is not intended for whole-body pediatric imaging. recalled?
Yes. FDA published recall Z-2040-2012 on August 1, 2012. Reported hazard: Please be aware that this is not a new recall. The firm has taken action; but, due to administrative issues this recall is now being classified by the Agency. The OEC 6800 Miniview was recalled in November 2006 due to several intermittent potential safety issues including navigation tracking inaccuracy, system Lock-up, incorrect image display, incorrect dose area calculated air kerma in "film mo
What should I do if I have this product?
Voluntary: Firm initiated · Letter