Leica M525 F50 Surgical Microscope; Article No. 10338422; Leica Microsystems (Schweiz) Ag, 201 Max Schmidheiny Strasse, Heerbrugg, Switzerland Product Usage: The intended use of this device is for improving visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.
Published August 1, 2012 · Recall initiated: June 4, 2012
Summary
FDA records medical device Class II recall Z-2053-2012, published August 1, 2012 and initiated June 4, 2012. Recalling firm: Leica Microsystems, Inc.; Status: Terminated; Reported quantity: 39 Units - USA , 4 Units - Canada; Lot: Serial numbers serial numbers 110311001, 120312001, 20412001, 260112001, 300911001, 40811001, 80411001, 90312001, 10311002, 90611001, 260711001, 20911001, 200911001, 210911001, 111011001, 141011001, 241011001, 121211001, 141211001, 151211001, 191211001, 30112001, 50112001, 140212001, 60312001, 220312001, 300312001, 30412001, 50411001, 10411001, 130411001, 210411001, 170511001, 300511001, 90811001, 200312001, 50412001, 110112002, 280611001, 241011002, 20312001, 101011001 and 270911001. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Leica Microsystems evaluated data showing that a malfunction resulting in an illumination reduction can be observed when Leica M525 F50 surgical microscope is used in combination with a remote control connected to a remote video adapter or motorized dual video adapter.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Leica Microsystems, Inc.
- Distributor
- Worldwide Distribution - US (nationwide) in the states of: California, Colorado, Florida, Georgia, Illinois, Indiana, Massachusetts, Michigan, Minnesota, New York, Oklahoma, Oregon, Pennsylvania, Texas, Utah, West Virginia, including Puerto Rico, and the countriy of Canada.
- Lot
- Serial numbers serial numbers 110311001, 120312001, 20412001, 260112001, 300911001, 40811001, 80411001, 90312001, 10311002, 90611001, 260711001, 20911001, 200911001, 210911001, 111011001, 141011001, 241011001, 121211001, 141211001, 151211001, 191211001, 30112001, 50112001, 140212001, 60312001, 220312001, 300312001, 30412001, 50411001, 10411001, 130411001, 210411001, 170511001, 300511001, 90811001, 200312001, 50412001, 110112002, 280611001, 241011002, 20312001, 101011001 and 270911001
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDABOND Ready-To-Use Primary Antibody CDX2 (EP25), REF PA0375September 22, 2021
- FDABOND Polymer Refine Detection Kit, Catalog No. DS9800. This detection system is for in vitro diagnostic use.December 18, 2019
- FDALeica M530 OHX Microscope System Supplied with 100 - 120 Vac.August 15, 2018
- FDANussloch GmbH ASP6025 Tissue ProcessorJuly 12, 2017
- FDAOEC 6800 Miniview, Image-intensified fluoroscopic x-ray system used during diagnostic, surgical andAugust 1, 2012
- FDAUniversal Spinal System Lamina Hooks Synthes Medium TI Lamina Hook - Right 498.320 Non sterile LOT 6876785August 1, 2012
Questions about this recall
Was Leica M525 F50 Surgical Microscope; Article No. 10338422; Leica Microsystems (Schweiz) Ag, 201 Max Schmidheiny Strasse, Heerbrugg, Switzerland Product Usage: The intended use of this device is for improving visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment. recalled?
Yes. FDA published recall Z-2053-2012 on August 1, 2012. Reported hazard: Leica Microsystems evaluated data showing that a malfunction resulting in an illumination reduction can be observed when Leica M525 F50 surgical microscope is used in combination with a remote control connected to a remote video adapter or motorized dual video adapter.
What should I do if I have this product?
Voluntary: Firm initiated · Letter