Leica M525 F50 Surgical Microscope; Article No. 10338422; Leica Microsystems (Schweiz) Ag, 201 Max Schmidheiny Strasse, Heerbrugg, Switzerland Product Usage: The intended use of this device is for improving visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment.

Published August 1, 2012 · Recall initiated: June 4, 2012

Summary

FDA records medical device Class II recall Z-2053-2012, published August 1, 2012 and initiated June 4, 2012. Recalling firm: Leica Microsystems, Inc.; Status: Terminated; Reported quantity: 39 Units - USA , 4 Units - Canada; Lot: Serial numbers serial numbers 110311001, 120312001, 20412001, 260112001, 300911001, 40811001, 80411001, 90312001, 10311002, 90611001, 260711001, 20911001, 200911001, 210911001, 111011001, 141011001, 241011001, 121211001, 141211001, 151211001, 191211001, 30112001, 50112001, 140212001, 60312001, 220312001, 300312001, 30412001, 50411001, 10411001, 130411001, 210411001, 170511001, 300511001, 90811001, 200312001, 50412001, 110112002, 280611001, 241011002, 20312001, 101011001 and 270911001. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Leica Microsystems evaluated data showing that a malfunction resulting in an illumination reduction can be observed when Leica M525 F50 surgical microscope is used in combination with a remote control connected to a remote video adapter or motorized dual video adapter.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Leica Microsystems, Inc.
Distributor
Worldwide Distribution - US (nationwide) in the states of: California, Colorado, Florida, Georgia, Illinois, Indiana, Massachusetts, Michigan, Minnesota, New York, Oklahoma, Oregon, Pennsylvania, Texas, Utah, West Virginia, including Puerto Rico, and the countriy of Canada.
Lot
Serial numbers serial numbers 110311001, 120312001, 20412001, 260112001, 300911001, 40811001, 80411001, 90312001, 10311002, 90611001, 260711001, 20911001, 200911001, 210911001, 111011001, 141011001, 241011001, 121211001, 141211001, 151211001, 191211001, 30112001, 50112001, 140212001, 60312001, 220312001, 300312001, 30412001, 50411001, 10411001, 130411001, 210411001, 170511001, 300511001, 90811001, 200312001, 50412001, 110112002, 280611001, 241011002, 20312001, 101011001 and 270911001
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Leica M525 F50 Surgical Microscope; Article No. 10338422; Leica Microsystems (Schweiz) Ag, 201 Max Schmidheiny Strasse, Heerbrugg, Switzerland Product Usage: The intended use of this device is for improving visibility of objects through magnification and illumination. It can be applied for observation and documentation and for human and veterinary medical treatment. recalled?

Yes. FDA published recall Z-2053-2012 on August 1, 2012. Reported hazard: Leica Microsystems evaluated data showing that a malfunction resulting in an illumination reduction can be observed when Leica M525 F50 surgical microscope is used in combination with a remote control connected to a remote video adapter or motorized dual video adapter.

What should I do if I have this product?

Voluntary: Firm initiated · Letter