Varian brand Varian Eclipse Treatment Planning System for radiotherapy; Model Number: H48; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye.
Published July 25, 2012 · Recall initiated: June 18, 2012
Summary
FDA records medical device Class II recall Z-2041-2012, published July 25, 2012 and initiated June 18, 2012. Recalling firm: Varian Medical Systems, Inc. Oncology Systems; Status: Terminated; Reported quantity: 7113 units; Lot: Affected Devices: H480110 ,H486186 ,H48B178 ,H48D026 ,H48D966 ,H487806 ,H48B764 ,H48B218 ,H48C503 ,H488467 ,H481815 ,H48D923 ,H48E486 ,H480267 ,H486446 ,H48B179 ,H48D027 ,H48A565 ,H487807 ,H48B765 ,H48B963 ,H48B756 ,H488468 ,H481816 ,H48E896 ,H48B426 ,H480268 ,H486447 ,H48B180 ,H48D028 ,H48A566 ,H487808 ,H482178 ,H48B964 ,H48B757 ,H489917 ,H485467 ,H48E897 ,H48B668 ,H480269 ,H487164 ,H48B181 ,H48D029 ,H48D566 ,H487809 ,H489962 ,H48B965 ,H482450 ,H48A912 ,H48A254 ,H489213 ,H48B669 ,H480270 ,H487165 ,H48B182 ,H48D030 ,H488558 ,H487810 ,H489963 ,H48B966 ,H482451 ,H48A913 ,H48C429 ,H489214 ,H48B670 ,H480271 ,H487166 ,H48B183 ,H48D031 ,H488559 ,H48D203 ,H48C022 ,H48C177 ,H482452 ,H48A914 ,H48B523 ,H489215 ,H48B671 ,H480272 ,H487167 ,H48B184 ,H48D032 ,H488560 ,H48D204 ,H48E169 ,H48C178 ,H483272 ,H48A915 ,H48B524 ,H484763 ,H48B672 ,H482174 ,H487168 ,H48B185 ,H48D033 ,H488561 ,H48D361 ,H48E170 ,H48C179 ,H484360 ,H48A916 ,H48B525 ,H484764 ,H48B673 ,H482934 ,H487169 ,H48B186 ,H481109 ,H489079 ,H48D362 ,H48E171 ,H485020 ,H488594 ,H48A917 ,H48B526 ,H484765 ,H48B674 ,H484123 ,H487170 ,H480953 ,H481110 ,H489110 ,H48D363 ,H48E172 ,H485021 ,H488595 ,H48C365 ,H48B527 ,H484766 ,H48B675 ,H485462 ,H487171 ,H480954 ,H481111 ,H48A614 ,H48D364 ,H48E493 ,H486023 ,H48A940 ,H48C366 ,H48B528 ,H485112 ,H48B676 ,H485590 ,H487172 ,H483209 ,H481323 ,H48A542 ,H48D365 ,H48E494 ,H488361 ,H48C141 ,H48E610 ,H489244 ,H486821 ,H48B677 ,H486422 ,H487173 ,H489760 ,H481671 ,H48D329 ,H48D366 ,H48E495 ,H488362 ,H48C142 ,H48E615 ,H489245 ,H489516 ,H48B678 ,H486423 ,H487974 ,H48E523 ,H481672 ,H48D330 ,H48D367 ,H489683 ,H488363 ,H48C143 ,H48E616 ,H489246 ,H489517 ,H48B679 ,H486424 ,H487975 ,H48E568 ,H481673 ,H48E385 ,H48D368 ,H489526 ,H488364 ,H48C144 ,H485992 ,H48A109 ,H489518 ,H48B680 ,H486425 ,H487976 ,H48E590 ,H481674 ,H481154 ,H480567 ,H489527 ,H48A703 ,H48C146 ,H485994 ,H48D432 ,H489519 ,H48B681 ,H486426 ,H48B150 ,H48E591 ,H481675 ,H481155 ,H480568 ,H489530 ,H48A704 ,H48C147 ,H485995 ,H48E389 ,H48C203 ,H48B682 ,H486427 ,H48B302 ,H48E592 ,H481676 ,H481156 ,H480569 ,H489531 ,H48A705 ,H489130 ,H485996 ,H482311 ,H48A274 ,H48B683 ,H486428 ,H48C019 ,H48E593 ,H484853 ,H483874 ,H480594 ,H48A253 ,H48A706 ,H489131 ,H48A952 ,H482312 ,H48A275 ,H48B684 ,H486429 ,H48C045 ,H480167 ,H484854 ,H489232 ,H480595 ,H48E088 ,H48A707 ,H489132 ,H48A953 ,H482313 ,H48A276 ,H48B685 ,H48B995 ,H48C046 ,H480168 ,H485093 ,H485656 ,H485633 ,H48E089 ,H48A708 ,H489133 ,H48A954 ,H482314 ,H48A277 ,H48B686 ,H48B996 ,H486633 ,H480285 ,H485094 ,H485657 ,H485634 ,H48E090 ,H48A709 ,H485678 ,H48A955 ,H482315 ,H48A278 ,H48B687 ,H481912 ,H486634 ,H481021 ,H485095 ,H485677 ,H485635 ,H48E091 ,H48D393 ,H485679 ,H48A960 ,H482316 ,H48B785 ,H48B688 ,H481913 ,H486635 ,H481608 ,H485096 ,H487661 ,H485636 ,H48E092 ,H48E317 ,H489094 ,H48A961 ,H482317 ,H48D894 ,H48B689 ,H481914 ,H486636 ,H482439 ,H485952 ,H487662 ,H485637 ,H48E093 ,H48E318 ,H48C294 ,H48A962 ,H482318 ,H48D895 ,H48B690 ,H481915 ,H486637 ,H483845 ,H485988 ,H487663 ,H485638 ,H48E094 ,H48E319 ,H48C295 ,H48A972 ,H482319 ,H48D896 ,H480732 ,H484396 ,H48C776 ,H48C412 ,H487126 ,H48B076 ,H485639 ,H48E095 ,H481927 ,H48C296 ,H48C290 ,H482320 ,H481309 ,H480748 ,H48D548 ,H483756 ,H48C413 ,H487127 ,H48B077 ,H485640 ,H48E096 ,H481928 ,H48E138 ,H48C291 ,H482321 ,H481310 ,H487708 ,H48D549 ,H488991 ,H48C414 ,H487128 ,H48D487 ,H485641 ,H489363 ,H481929 ,H48E140 ,H48C292 ,H482322 ,H482287 ,H487709 ,H48D550 ,H488675 ,H48C415 ,H487129 ,H48D488 ,H486017 ,H489364 ,H481930 ,H48E144 ,H48C293 ,H484876 ,H483517 ,H489558 ,H48D551 ,H48C047 ,H48C416 ,H487640 ,H480329 ,H48D340 ,H489365 ,H481931 ,H48E146 ,H48B194 ,H484877 ,H484398 ,H489559 ,H481551 ,H48C048 ,H486509 ,H487641 ,H480339 ,H48D341 ,H489366 ,H489856 ,H48E336 ,H48B195 ,H484878 ,H486308 ,H48A941 ,H482656 ,H48C049 ,H48D983 ,H487642 ,H480960 ,H48D342 ,H489418 ,H489857 ,H48E337 ,H48C975 ,H484879 ,H486311 ,H480378 ,H482657 ,H48E524 ,H481215 ,H48B153 ,H481640 ,H48D347 ,H488582 ,H489858 ,H48E338 ,H48. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Varian has identified an anomaly with the Eclipse Treatment Planning (for radiotherapy treatments) where changing the prescription in a calculated VMAT or RapidArc plan may lead to Monitor Units which do not reflect the new prescription. If the dose prescription of a VMAT or RapidArc plan is changed after the initial Dose Volume Optimization, the Monitor Units and dose distribution may not reflect
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Varian Medical Systems, Inc. Oncology Systems
- Distributor
- Worldwide distribution: USA (nationwide) and countries including of: Austria, Australia, Bahrain, Belgium, Brazil, Canada, China, Chile, Colombia, Denmark, Egypt, Finland, France,Germany, Greece, Hungary, Hong Kong, India, Indonesia, Israel, Ireland, Italy, Japan, Latvia, Malaysia, Mexico, Morocco, Nepal, Netherlands, Norway, Panama, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, United Kingdom, Taiwan, Turkey, Ukraine, United Arab Emirates and Venezuela.
- Lot
- Affected Devices: H480110 ,H486186 ,H48B178 ,H48D026 ,H48D966 ,H487806 ,H48B764 ,H48B218 ,H48C503 ,H488467 ,H481815 ,H48D923 ,H48E486 ,H480267 ,H486446 ,H48B179 ,H48D027 ,H48A565 ,H487807 ,H48B765 ,H48B963 ,H48B756 ,H488468 ,H481816 ,H48E896 ,H48B426 ,H480268 ,H486447 ,H48B180 ,H48D028 ,H48A566 ,H487808 ,H482178 ,H48B964 ,H48B757 ,H489917 ,H485467 ,H48E897 ,H48B668 ,H480269 ,H487164 ,H48B181 ,H48D029 ,H48D566 ,H487809 ,H489962 ,H48B965 ,H482450 ,H48A912 ,H48A254 ,H489213 ,H48B669 ,H480270 ,H487165 ,H48B182 ,H48D030 ,H488558 ,H487810 ,H489963 ,H48B966 ,H482451 ,H48A913 ,H48C429 ,H489214 ,H48B670 ,H480271 ,H487166 ,H48B183 ,H48D031 ,H488559 ,H48D203 ,H48C022 ,H48C177 ,H482452 ,H48A914 ,H48B523 ,H489215 ,H48B671 ,H480272 ,H487167 ,H48B184 ,H48D032 ,H488560 ,H48D204 ,H48E169 ,H48C178 ,H483272 ,H48A915 ,H48B524 ,H484763 ,H48B672 ,H482174 ,H487168 ,H48B185 ,H48D033 ,H488561 ,H48D361 ,H48E170 ,H48C179 ,H484360 ,H48A916 ,H48B525 ,H484764 ,H48B673 ,H482934 ,H487169 ,H48B186 ,H481109 ,H489079 ,H48D362 ,H48E171 ,H485020 ,H488594 ,H48A917 ,H48B526 ,H484765 ,H48B674 ,H484123 ,H487170 ,H480953 ,H481110 ,H489110 ,H48D363 ,H48E172 ,H485021 ,H488595 ,H48C365 ,H48B527 ,H484766 ,H48B675 ,H485462 ,H487171 ,H480954 ,H481111 ,H48A614 ,H48D364 ,H48E493 ,H486023 ,H48A940 ,H48C366 ,H48B528 ,H485112 ,H48B676 ,H485590 ,H487172 ,H483209 ,H481323 ,H48A542 ,H48D365 ,H48E494 ,H488361 ,H48C141 ,H48E610 ,H489244 ,H486821 ,H48B677 ,H486422 ,H487173 ,H489760 ,H481671 ,H48D329 ,H48D366 ,H48E495 ,H488362 ,H48C142 ,H48E615 ,H489245 ,H489516 ,H48B678 ,H486423 ,H487974 ,H48E523 ,H481672 ,H48D330 ,H48D367 ,H489683 ,H488363 ,H48C143 ,H48E616 ,H489246 ,H489517 ,H48B679 ,H486424 ,H487975 ,H48E568 ,H481673 ,H48E385 ,H48D368 ,H489526 ,H488364 ,H48C144 ,H485992 ,H48A109 ,H489518 ,H48B680 ,H486425 ,H487976 ,H48E590 ,H481674 ,H481154 ,H480567 ,H489527 ,H48A703 ,H48C146 ,H485994 ,H48D432 ,H489519 ,H48B681 ,H486426 ,H48B150 ,H48E591 ,H481675 ,H481155 ,H480568 ,H489530 ,H48A704 ,H48C147 ,H485995 ,H48E389 ,H48C203 ,H48B682 ,H486427 ,H48B302 ,H48E592 ,H481676 ,H481156 ,H480569 ,H489531 ,H48A705 ,H489130 ,H485996 ,H482311 ,H48A274 ,H48B683 ,H486428 ,H48C019 ,H48E593 ,H484853 ,H483874 ,H480594 ,H48A253 ,H48A706 ,H489131 ,H48A952 ,H482312 ,H48A275 ,H48B684 ,H486429 ,H48C045 ,H480167 ,H484854 ,H489232 ,H480595 ,H48E088 ,H48A707 ,H489132 ,H48A953 ,H482313 ,H48A276 ,H48B685 ,H48B995 ,H48C046 ,H480168 ,H485093 ,H485656 ,H485633 ,H48E089 ,H48A708 ,H489133 ,H48A954 ,H482314 ,H48A277 ,H48B686 ,H48B996 ,H486633 ,H480285 ,H485094 ,H485657 ,H485634 ,H48E090 ,H48A709 ,H485678 ,H48A955 ,H482315 ,H48A278 ,H48B687 ,H481912 ,H486634 ,H481021 ,H485095 ,H485677 ,H485635 ,H48E091 ,H48D393 ,H485679 ,H48A960 ,H482316 ,H48B785 ,H48B688 ,H481913 ,H486635 ,H481608 ,H485096 ,H487661 ,H485636 ,H48E092 ,H48E317 ,H489094 ,H48A961 ,H482317 ,H48D894 ,H48B689 ,H481914 ,H486636 ,H482439 ,H485952 ,H487662 ,H485637 ,H48E093 ,H48E318 ,H48C294 ,H48A962 ,H482318 ,H48D895 ,H48B690 ,H481915 ,H486637 ,H483845 ,H485988 ,H487663 ,H485638 ,H48E094 ,H48E319 ,H48C295 ,H48A972 ,H482319 ,H48D896 ,H480732 ,H484396 ,H48C776 ,H48C412 ,H487126 ,H48B076 ,H485639 ,H48E095 ,H481927 ,H48C296 ,H48C290 ,H482320 ,H481309 ,H480748 ,H48D548 ,H483756 ,H48C413 ,H487127 ,H48B077 ,H485640 ,H48E096 ,H481928 ,H48E138 ,H48C291 ,H482321 ,H481310 ,H487708 ,H48D549 ,H488991 ,H48C414 ,H487128 ,H48D487 ,H485641 ,H489363 ,H481929 ,H48E140 ,H48C292 ,H482322 ,H482287 ,H487709 ,H48D550 ,H488675 ,H48C415 ,H487129 ,H48D488 ,H486017 ,H489364 ,H481930 ,H48E144 ,H48C293 ,H484876 ,H483517 ,H489558 ,H48D551 ,H48C047 ,H48C416 ,H487640 ,H480329 ,H48D340 ,H489365 ,H481931 ,H48E146 ,H48B194 ,H484877 ,H484398 ,H489559 ,H481551 ,H48C048 ,H486509 ,H487641 ,H480339 ,H48D341 ,H489366 ,H489856 ,H48E336 ,H48B195 ,H484878 ,H486308 ,H48A941 ,H482656 ,H48C049 ,H48D983 ,H487642 ,H480960 ,H48D342 ,H489418 ,H489857 ,H48E337 ,H48C975 ,H484879 ,H486311 ,H480378 ,H482657 ,H48E524 ,H481215 ,H48B153 ,H481640 ,H48D347 ,H488582 ,H489858 ,H48E338 ,H48
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVarian High Energy Clinacs, High Energy Accelerator, Radiation Treatment System, Model Numbers: H14, H27, H29,May 22, 2013
- FDAVarian brand RPM Respiratory Gating System, RPM Gating v1.7.5 with 3D Option, Model Number: HS2 Product Usage:May 1, 2013
- FDAVarian Eclipse Treatment Planning System for Radiotherapy; Model Number: H48. In addition, the Eclipse ProtonFebruary 27, 2013
- FDAVarian brand ARIA Radiation Oncology, ARIA Oncology Information System Radiation Oncology, Import ExportDecember 26, 2012
- FDAAlere Triage Tox+MTD Drug Screen, PN 94400 Product Usage: The Alere Triage TOX Drug Screen is a fluorescenceJuly 25, 2012
- FDATerumo Advanced Perfusion System 1. System 1 base 100/120V. The Terumo Advanced Perfusion System I isJuly 25, 2012
Questions about this recall
Was Varian brand Varian Eclipse Treatment Planning System for radiotherapy; Model Number: H48; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The Eclipse Treatment Planning System (Eclipse TPS) is used to plan radiotherapy treatments for patients with malignant or benign diseases. Eclipse TPS is used to plan external beam irradiation with photon, electron and proton beams, as well as for internal irradiation (brachytherapy) treatments. In addition, the Eclipse Proton Eye algorithm is specifically indicated for planning proton treatment of neoplasms of the eye. recalled?
Yes. FDA published recall Z-2041-2012 on July 25, 2012. Reported hazard: Varian has identified an anomaly with the Eclipse Treatment Planning (for radiotherapy treatments) where changing the prescription in a calculated VMAT or RapidArc plan may lead to Monitor Units which do not reflect the new prescription. If the dose prescription of a VMAT or RapidArc plan is changed after the initial Dose Volume Optimization, the Monitor Units and dose distribution may not reflect
What should I do if I have this product?
Voluntary: Firm initiated · Letter