FDAFDA · Food and drugs·Recall number Z-2164-2013
Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primary and secondary fluids when positive pressure is required. The infusion system is capable of delivering fluid from a negative head height (when the IV fluid container is lower than the pump), and provides clinically accepted volumetric accuracy for all standard IV fluids, including blood, lipid, and Total Parenteral Nutrition (TPN). Model number: 621-100ES, 621-200ES, 621-300ES, and 621-400ES. To deliver a broad range of drugs with use with the B. Braun Medical Inc. Horizon Pump IV Sets.
Published September 18, 2013 · Recall initiated: October 31, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
This notice is being provided in follow up to the field correction initiated by B. Braun Medical Inc. on 26 August, 2011 due to an issue in the Outlook ES Safety Infusion System where the Care Area can inadvertently be exited thereby removing DoseGuard dose limit protection. This occurs when a sequence of HOLD - HOLD key strokes are performed on the pump key panel while the user is in the Dose Dat
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- B Braun Medical, Inc.
- Distributor
- Nationwide Distribution.
- Lot
- Model number: 621-100ES, 621-200ES, 621-300ES, and 621-400ES with main processor software versions prior to 151580. Multiple serial numbers affected. Affected serial numbers were notified of this field correction.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States