Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primary and secondary fluids when positive pressure is required. The infusion system is capable of delivering fluid from a negative head height (when the IV fluid container is lower than the pump), and provides clinically accepted volumetric accuracy for all standard IV fluids, including blood, lipid, and Total Parenteral Nutrition (TPN). Model number: 621-300ES and 621-400ES. To deliver a broad range of drugs with use with the B. Braun Medical Inc. Horizon Pump IV Sets.
Published September 25, 2013 · Recall initiated: November 14, 2012
Summary
FDA records medical device Class I recall Z-2188-2013, published September 25, 2013 and initiated November 14, 2012. Recalling firm: B Braun Medical, Inc.; Status: Terminated; Reported quantity: 6,919; Lot: Model number: 621-300ES and 621-400ES. Multiple serial numbers affected. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
This notice is being provided in follow up to the field correction initiated by B. Braun Medical Inc. on August 26, 2011 due to an issue in which the Outlook ES Safety Infusion System may halt infusion but the "RUN" Light Emitting Diodes (LEDs) on the front display continue to advance as if the pump were infusing. The pump emits a backup alarm, but there are no visual indicators that the infusion
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- B Braun Medical, Inc.
- Distributor
- Nationwide distribution: USA including states of: AR, CA, FL, IL, LA, NC, NJ, NY, MD, ME, MN, SC, TN, TX, VA, WI, and WV.
- Lot
- Model number: 621-300ES and 621-400ES. Multiple serial numbers affected.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAPEX Compounding System Control Panel Module, REF: 601237, Product Code AX1000 - Product Usage: intended forDecember 30, 2020
- FDAJuno DRF System. Mobile X-ray system These are general, multi-function x-ray systems capable of routineSeptember 25, 2013
- FDADiaScreen Liquid Urine Controls: Positive and Negative Assayed Liquid Quality Controls for use with DipsticksSeptember 25, 2013
- FDAOutlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flowSeptember 18, 2013
- FDAOutlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flowSeptember 18, 2013
- FDAOutlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flowSeptember 18, 2013
Questions about this recall
Was Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primary and secondary fluids when positive pressure is required. The infusion system is capable of delivering fluid from a negative head height (when the IV fluid container is lower than the pump), and provides clinically accepted volumetric accuracy for all standard IV fluids, including blood, lipid, and Total Parenteral Nutrition (TPN). Model number: 621-300ES and 621-400ES. To deliver a broad range of drugs with use with the B. Braun Medical Inc. Horizon Pump IV Sets. recalled?
Yes. FDA published recall Z-2188-2013 on September 25, 2013. Reported hazard: This notice is being provided in follow up to the field correction initiated by B. Braun Medical Inc. on August 26, 2011 due to an issue in which the Outlook ES Safety Infusion System may halt infusion but the "RUN" Light Emitting Diodes (LEDs) on the front display continue to advance as if the pump were infusing. The pump emits a backup alarm, but there are no visual indicators that the infusion
What should I do if I have this product?
Voluntary: Firm initiated · Letter