Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primary and secondary fluids when positive pressure is required. The infusion system is capable of delivering fluid from a negative head height (when the IV fluid container is lower than the pump), and provides clinically accepted volumetric accuracy for all standard IV fluids, including blood, lipid, and Total Parenteral Nutrition (TPN). Model number: 621-100ES, and 621-200ES. To deliver a broad range of drugs with use with the B. Braun Medical Inc. Horizon Pump IV Sets.
Published September 18, 2013 · Recall initiated: December 5, 2012
Summary
FDA records medical device Class II recall Z-2175-2013, published September 18, 2013 and initiated December 5, 2012. Recalling firm: B Braun Medical, Inc.; Status: Terminated; Reported quantity: 14,161; Lot: Models 620-100 and 620-200; multiple serial numbers. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
B. Braun Medical Inc. is informing our customers of a field correction previously performed to address the loss of the drug library in the Outlook Pump Models 620-100 and 620-200. This could have occurred after periods of lost DC (battery) power or out of tolerance low power. This could have led to a delay in therapy, or to delivery of an over or under-dose of a therapy if the user was to rely on
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- B Braun Medical, Inc.
- Distributor
- Worldwide Disribution: USA (nationwide) and country of: Canada.
- Lot
- Models 620-100 and 620-200; multiple serial numbers
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAPEX Compounding System Control Panel Module, REF: 601237, Product Code AX1000 - Product Usage: intended forDecember 30, 2020
- FDAOutlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flowSeptember 25, 2013
- FDAOutlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flowSeptember 18, 2013
- FDAOutlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flowSeptember 18, 2013
- FDASTERRAD 100NX Cassettes, Product Code: 10144 The STERRAD Sterilization System is a low temperature generalSeptember 18, 2013
- FDASurgical microscope The device is used to improve the visibility of objects through magnification andSeptember 18, 2013
Questions about this recall
Was Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primary and secondary fluids when positive pressure is required. The infusion system is capable of delivering fluid from a negative head height (when the IV fluid container is lower than the pump), and provides clinically accepted volumetric accuracy for all standard IV fluids, including blood, lipid, and Total Parenteral Nutrition (TPN). Model number: 621-100ES, and 621-200ES. To deliver a broad range of drugs with use with the B. Braun Medical Inc. Horizon Pump IV Sets. recalled?
Yes. FDA published recall Z-2175-2013 on September 18, 2013. Reported hazard: B. Braun Medical Inc. is informing our customers of a field correction previously performed to address the loss of the drug library in the Outlook Pump Models 620-100 and 620-200. This could have occurred after periods of lost DC (battery) power or out of tolerance low power. This could have led to a delay in therapy, or to delivery of an over or under-dose of a therapy if the user was to rely on
What should I do if I have this product?
Voluntary: Firm initiated · Letter