Back to search
FDAFDA · Food and drugs·Recall number Z-2183-2016

Target Detachable Coils: TARGET HELICAL NANO 2.5MM X 4CM MODEL Number:M0035452540 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.

Published July 20, 2016 · Recall initiated: June 6, 2016

Summary

FDA records medical device Class II recall Z-2183-2016, published July 20, 2016 and initiated June 6, 2016. Recalling firm: Stryker Neurovascular; Status: Terminated; Reported quantity: 45,302 (8,188 in the U.S.) - total, all model numbers; Lot: MODEL Number:M0035452540; Lot Number/Expiration Date: 18440351 30-Jun-18 18539018 31-Aug-18 18439805 30-Jun-18 18519562 31-Jul-18 18468754 30-Jun-18 18922401 31-Mar-19 19060307 31-May-19 18970593 30-Apr-19 18898116 28-Feb-19 19034354 31-May-19 18875384 28-Feb-19. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Stryker Neurovascular
Distributor
Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.
Lot
MODEL Number:M0035452540; Lot Number/Expiration Date: 18440351 30-Jun-18 18539018 31-Aug-18 18439805 30-Jun-18 18519562 31-Jul-18 18468754 30-Jun-18 18922401 31-Mar-19 19060307 31-May-19 18970593 30-Apr-19 18898116 28-Feb-19 19034354 31-May-19 18875384 28-Feb-19.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Target Detachable Coils: TARGET HELICAL NANO 2.5MM X 4CM MODEL Number:M0035452540 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels. recalled?

Yes. FDA published recall Z-2183-2016 on July 20, 2016. Reported hazard: Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.

What should I do if I have this product?

Voluntary: Firm initiated · Letter