Target Detachable Coils: TARGET HELICAL NANO 3MM X 4CM MODEL Number:M0035453040 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels.
Published July 20, 2016 · Recall initiated: June 6, 2016
Summary
FDA records medical device Class II recall Z-2185-2016, published July 20, 2016 and initiated June 6, 2016. Recalling firm: Stryker Neurovascular; Status: Terminated; Reported quantity: 45,302 (8,188 in the U.S.) - total, all model numbers; Lot: MODEL Number:M0035453040; Lot Number/Expiration Date: 18603553 30-Sep-18. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Neurovascular
- Distributor
- Worldwide Distribution - US (nationwide) and Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Egypt, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Italia, Japan,Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, Luxembourg, Mexico,Netherlands, Norway, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Turkey, United Arab Emirates, United Kingdom, United States, Uruguay, and Vietnam.
- Lot
- MODEL Number:M0035453040; Lot Number/Expiration Date: 18603553 30-Sep-18
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAINZONE DETACHMENT SYSTEM, REF: M00345100950July 1, 2026
- FDAThe Trevo Trak 21 Microcatheter is a single-lumen, braided shaft, variable stiffness catheter with radiopaqueOctober 16, 2024
- FDAThe Synchro SELECT Guidewires are a steerable guidewire family with a shapeable tip and are available inJune 12, 2024
- FDAThe Synchro Neuro Guidewires are a steerable guidewire family with a shapeable tip. The outside diameter ofJune 12, 2024
- FDATarget Detachable Coils: TARGET HELICAL NANO 2.5MM X 4CM MODEL Number:M0035452540 Neurology: Target DetachableJuly 20, 2016
- FDAPart # 313.232, 2.0mm Stardrive Screwdriver Bld Slf-Retain Med/66mm MQC, lot # 4363820, mfg. 9/11/2002 ProductJuly 20, 2016
Questions about this recall
Was Target Detachable Coils: TARGET HELICAL NANO 3MM X 4CM MODEL Number:M0035453040 Neurology: Target Detachable Coils are intended to endovascularly obstruct or occlude blood flow in vascular abnormalities of the neurovascular and peripheral vessels. recalled?
Yes. FDA published recall Z-2185-2016 on July 20, 2016. Reported hazard: Stryker Neurovascular has become aware that some units of Target Nano product do not have the intended stretch resistance. The sutures inside the coil that provide stretch resistance may have been damaged during manufacturing which can lead to an increased occurrence of coil stretching.
What should I do if I have this product?
Voluntary: Firm initiated · Letter