FDAFDA · Food and drugs·Recall number Z-2187-2021

REF 8L92-42, Abbott Activated Alanine Aminotransferase, for use with Architect, IVD, CE, UDI: (01) 00380740161866

Published August 11, 2021 · Recall initiated: June 10, 2021

Summary

FDA records medical device Class II recall Z-2187-2021, published August 11, 2021 and initiated June 10, 2021. Recalling firm: Abbott Laboratories; Status: Ongoing; Reported quantity: 2,212 units; Lot: Lot Numbers / Expiration Date 84126UN20 / 09JUL2021 03676UN20 / 25AUG2021 10597UN20 / 29AUG2021 26275UN20 / 20OCT2021 77748UN20 / 08DEC2021 65958UN20 / 08DEC2021 81824UN21 / 16FEB2022. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a potential for >10% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Abbott Laboratories
Distributor
Worldwide distribution - US Nationwide distribution in the states of VA, UT, TX, TN, SD,PA, NY, ME, ID, CA and the countries of AFGHANISTAN, AUSTRALIA, AUSTRIA, BAHAMAS, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVI, CANADA, CAYMAN ISLANDS, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, FAROE ISLAND, FINLAND, FRANCE, GERMANY, GREENLAND, Hong Kong, INDIA, IRELAND, ISRAEL, ITALY, KUWAIT, LITHUANIA, LUXEMBOURG, MEXICO, MONTENEGRO, NETHERLANDS, NORWAY, OMAN, PAKISTAN, PANAMA, PHILIPPINES, PORTUGAL, PR CHINA, REPUBLIC OF YEMEN, ROMANIA, RUSSIA, SAUDI ARABIA, SLOVAKIA, SLOVENIA, SPAIN, SWEDEN, SWITZERLAND, THAILAND, TRINIDAD & TOBAGO, UNITED KINGDOM, VIETNAM.
Lot
Lot Numbers / Expiration Date 84126UN20 / 09JUL2021 03676UN20 / 25AUG2021 10597UN20 / 29AUG2021 26275UN20 / 20OCT2021 77748UN20 / 08DEC2021 65958UN20 / 08DEC2021 81824UN21 / 16FEB2022
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was REF 8L92-42, Abbott Activated Alanine Aminotransferase, for use with Architect, IVD, CE, UDI: (01) 00380740161866 recalled?

Yes. FDA published recall Z-2187-2021 on August 11, 2021. Reported hazard: There is a potential for >10% under-recovery on Activated Alanine Aminotransferase (A-ALT) samples greater than 1,200 U/L. In addition there is a potential for a >10% shift in patient results with samples containing bilirubin concentrations greater that 31 mg/dL. This could lead to incorrect patient results

What should I do if I have this product?

Voluntary: Firm initiated · Letter