COBALT XT HF QUAD CRT-D MRI SureScan, Model Number DTPA2Q1, Implantable Cardioverter Defibrillator

Published July 3, 2024 · Recall initiated: April 16, 2024

Summary

FDA records medical device Class II recall Z-2193-2024, published July 3, 2024 and initiated April 16, 2024. Recalling firm: Medtronic Inc.; Status: Ongoing; Reported quantity: 11 units; Lot: GTIN: 00763000178109, Serial Numbers: RTD607545S, RTD607549S, RTD607551S; GTIN 00763000711177, Serial Numbers: RTD607363S, RTD607364S, RTD607367S, RTD607369S, RTD607370S, RTD607374S, RTD607404S, RTD607405S. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing.

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Medtronic Inc.
Distributor
Worldwide - US Nationwide distribution in the states of CA, CT, FL, IN, MA, MI, NC, NH, NJ, NM, NY, OH, TX, VA and the countries of Australia, Austria, Bulgaria, Canada, China, Croatia, Czech Republic, France, Georgia, Germany, Greece, Guadeloupe, Iceland, Ireland, Italy, Jordan, Kosovo, Latvia, Moldova, Republic Of, Netherlands, Northern Ireland, Norway, Oman, Panama, Poland, Portugal, Qatar, Reunion, Slovakia, Slovenia, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
Lot
GTIN: 00763000178109, Serial Numbers: RTD607545S, RTD607549S, RTD607551S; GTIN 00763000711177, Serial Numbers: RTD607363S, RTD607364S, RTD607367S, RTD607369S, RTD607370S, RTD607374S, RTD607404S, RTD607405S.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was COBALT XT HF QUAD CRT-D MRI SureScan, Model Number DTPA2Q1, Implantable Cardioverter Defibrillator recalled?

Yes. FDA published recall Z-2193-2024 on July 3, 2024. Reported hazard: Potential for manufacturing defect. Weld crack led to devices failing at the leak check step in manufacturing.

What should I do if I have this product?

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit