STAINLESS STEEL REPROCESSING TRAY, 8MM ENDOSCOPE, IS4000, Model No. 400498-02 - Product Usage: is designed to enable complex surgery using a minimally invasive approach.

Published June 17, 2020 · Recall initiated: December 19, 2019

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Small but detectable holes in the sterilization wrap used with the firm's 8 mm endoscope sterilization trays and handheld camera sterilization trays could lead to a potential breach of sterility and risk of infection. Certain features on the trays with handling and storage may cause these holes.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Intuitive Surgical, Inc.
Distributor
Worldwide distribution - US Nationwide distribution including in the states of Arkansas, Arizona, Alabama, Alaska, California, Connecticut, Colorado, Florida, Georgia, Hawaii, Illinois, Indiana, Iowa, Idaho, Kentucky, Kansas, Louisiana, Missouri, Michigan, Maryland, Mississippi, Minnesota, Montana, Maine, Massachusetts, Nevada, New York, North Dakota, Nebraska, North Carolina, New Jersey, Ohio, Oklahoma, Oregon, Pennsylvania, Rhode Island, South Carolina, Tennessee, Texas, Utah, Virginia, Wisconsin, Washington and West Virginia. Multiple countries including Canada.
Lot
Model No. 400498-02; All Batch Numbers; UDI: 00886874115848
Risk classification
Class II
Area (FDA)
device
Country
United States