Mazor X Surgical System Positioner Type II, REF: ASM0214-02

Published June 17, 2020 · Recall initiated: December 5, 2019

Summary

FDA records medical device Class II recall Z-2330-2020, published June 17, 2020 and initiated December 5, 2019. Recalling firm: Medtronic Navigation, Inc.; Status: Ongoing; Reported quantity: 177 Systems; Lot: All Lots. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is a potential for the surgical system to detach from the operational room table unexpectedly. This could result in the surgical system falling onto the patient resulting in blunt injury [hemorrhage, hematoma, bleeding, or fracture(s)]

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Medtronic Navigation, Inc.
Distributor
US: ALAR,AZ,CA,CO,CT,FL,GA,IA,ID,IL,IN,KY,MA,MI,MN,MO,NC,ND,NL,NY,OH,PA,SC,TN,TX,UT,VA,WA,WI,WV, OUS: Australia, China, India, Israel, Romania, Singapore, Thailand, United Kingdom, Switzerland, Spain, Germany, Belgium,
Lot
All Lots
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Mazor X Surgical System Positioner Type II, REF: ASM0214-02 recalled?

Yes. FDA published recall Z-2330-2020 on June 17, 2020. Reported hazard: There is a potential for the surgical system to detach from the operational room table unexpectedly. This could result in the surgical system falling onto the patient resulting in blunt injury [hemorrhage, hematoma, bleeding, or fracture(s)]

What should I do if I have this product?

Voluntary: Firm initiated · Letter