RayStation 5 Service Pack 3, UDI # 07350002010020 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.

Published June 17, 2020 · Recall initiated: May 18, 2020

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.

What to do

Voluntary: Firm initiated · E-Mail

Recall details

Manufacturer
RAYSEARCH LABORATORIES AB
Distributor
US Nationwide distribution including in the states of Florida, Georgia, Utah, Virginia, California, Maryland, Arizona, Texas, Washington DC, Missouri, Massachusetts, Illinois, Oklahoma, New Jersey, Tennessee, and Louisiana.
Lot
UDI # 07350002010020 Build number: 5.0.3.17
Risk classification
Class II
Area (FDA)
device
Country
Sweden