RayStation 6 Service Pack 2, UDI # 07350002010075 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.
Published June 17, 2020 · Recall initiated: May 18, 2020
Summary
FDA records medical device Class II recall Z-2342-2020, published June 17, 2020 and initiated May 18, 2020. Recalling firm: RAYSEARCH LABORATORIES AB; Status: Terminated; Reported quantity: N/A; Lot: UDI # 07350002010075 Build number: 6.2.0.7. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- RAYSEARCH LABORATORIES AB
- Distributor
- US Nationwide distribution including in the states of Florida, Georgia, Utah, Virginia, California, Maryland, Arizona, Texas, Washington DC, Missouri, Massachusetts, Illinois, Oklahoma, New Jersey, Tennessee, and Louisiana.
- Lot
- UDI # 07350002010075 Build number: 6.2.0.7
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Sweden
Related recalls
- FDARayCare 2024A SP4. Oncology Information Systems.July 8, 2026
- FDARayCare 2024A SP1. Oncology Information Systems.July 8, 2026
- FDABrand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 14.0.0, 15.0.0, 15.1.3 SoftwareJanuary 28, 2026
- FDABrand Name: RayStation Product Name: RayStation/RayPlan Model/Catalog Number: 17.0.0, 17.0.1 Software Version:January 28, 2026
- FDARayStation 5 Service Pack 3, UDI # 07350002010020 - Product Usage: For radiation therapy and medical oncologyJune 17, 2020
- FDARayStation 6 Service Pack 3, UDI # 07350002010242 - Product Usage: For radiation therapy and medical oncologyJune 17, 2020
Questions about this recall
Was RayStation 6 Service Pack 2, UDI # 07350002010075 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments. recalled?
Yes. FDA published recall Z-2342-2020 on June 17, 2020. Reported hazard: Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail